Article Abstract

Objective: During the advent of laparoscopy, surgeons directly explored the abdominal cavity with a telescope-like device through a small incision. Since then, numerous technological advances have transformed minimally invasive surgery (MIS). Yet, in our wireless world, various devices crowd the surgical field, with long wires and light sources posing fall and fire risks. The primary objective of this study was to analyze the first reported human use of a novel wireless laparoscopy system or WLS (ArthroFree™, Lazurite®, Cleveland, Ohio).

Methods: The utility and convenience of the WLS was assessed via two avenues: (1) by analyzing surgical outcomes from first human use and (2) by surveying healthcare professionals regarding its quality and utility.

Results: Eighteen patients (mean age 44.2, 83.3% female, mean BMI 33.4) underwent operations with the WLS. Operations included gynecologic and general surgical procedures. There were no intraoperative or postoperative complications, and no conversions to traditional laparoscopy or laparotomy. Mean operating time was 71.94 ± 20.41 min, and estimated blood loss was minimal. Survey results revealed varied individual experiences. Strengths included adequate illumination, improved ergonomics, and simplicity of setup and ease of operation. One respondent criticized the image resolution. Feedback indicated an overall positive impact, and 67% of respondents supported inclusion of the device at their facility. Moreover, its deployment in resource-limited settings abroad has demonstrated its efficacy in global surgery, indicating its potential in various healthcare environments.

Conclusions: This is the first reported human use of a novel WLS. Clinical results supported efficiency and safety of the technology. The successful deployment of the WLS in diverse surgical environments, including resource-limited settings, highlights its potential as a universally adaptable tool in global surgery. This report represents a strong first step toward a wireless operating room with the promise of redefining surgical standards as well as bridging gaps in surgical care worldwide.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11525237PMC
http://dx.doi.org/10.1007/s00464-024-11286-9DOI Listing

Publication Analysis

Top Keywords

reported human
12
human novel
8
resource-limited settings
8
global surgery
8
surgical
6
wls
5
wireless
4
human wireless
4
wireless laparoscopic
4
laparoscopic system
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!