Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Introduction: International guidelines suggest different possibilities for drying of endoscopes during reprocessing. Clinical results of these available drying methods are not satisfactory. The aim of this study was to compare the drying cycle of a standard endoscope washer-disinfector (EWD) (standard drying method [SD]) with a shortened mandatory drying by the EWD followed by a special drying device using laminar and turbulent air flow (novel drying method [ND]).
Methods: Sixty endoscopes (duodenoscopes, colonoscocopes, and gastroscopes) from 3 different manufacturers underwent high-level disinfection and drying depending on the randomization group. Operational time of drying was measured for both groups. Residual fluid in the channels was measured using a laboratory scale. After a 14-day storage period, a sample of the endoscope channels was obtained to determine bacterial contamination.
Results: ND had significantly fewer residual water in endoscope channels (SD: 90% vs ND: 0%; P < 0.001) after high-level disinfection and drying and less bacterial contamination after storage for 14 days (SD: 47% vs ND: 20%; P = 0.028). Time consumed for drying in ND was also significantly shorter (SD: 16 minutes 4 seconds vs ND: 5 minutes 59 seconds; P < 0.001).
Discussion: Drying with a special automatic drying device was superior compared with an EWD's drying program as evidenced by no measurable residual water, reduced microbiological contamination, and a more than 2-fold decrease in operational time. Thus, drying by laminar and turbulent airflow may represent an attractive alternative to the currently used standard approach in the reprocessing process of flexible endoscopes.