Background And Purpose: Monte Carlo (MC) based dose calculations are widely used in radiotherapy with a low statistical uncertainty, being accurate but slow. Increasing the uncertainty accelerates the calculation, but reduces quality. In online adaptive planning, however, dose is recalculated every treatment fraction, potentially decreasing the cumulative calculation error. This study aimed to evaluate the effect of higher MC statistical uncertainty in the context of daily online plan adaptation.

Materials And Methods: For twenty prostate cancer patients, daily plans were simulated for 5 fractions and three modes of variation: rigid whole body translations, local-rigid prostate translations and local-rigid prostate rotations. For each mode and fraction, adaptive plans were generated from a clinical reference plan using three MC uncertainty values: 1 % (standard), 2 % and 3 % per plan. Dose-volume criteria were evaluated for accumulated doses, checking plan acceptability and comparing higher uncertainty plans to the standard.

Results: Increasing the statistical uncertainty setting from 1 % to 2-3 % caused an accumulated median target D reduction of 0.1 Gy, with interquartile ranges (IQRs) up to 0.12 Gy. Rectum V increased in median up to 0.16 cm with IQRs up to 0.33 cm. The bladder V and V showed median increases up to 0.24 %-point, with IQRs up to 0.54 %-point. Using 2 % uncertainty reduced calculation times by more than a minute for all modes of variation, with no further time gain when increasing to 3 %.

Conclusion: A 2-3 % MC statistical uncertainty was clinically feasible. Using a 2 % uncertainty setting reduced calculation times at the cost of limited relative dose-volume differences.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11405816PMC
http://dx.doi.org/10.1016/j.phro.2024.100636DOI Listing

Publication Analysis

Top Keywords

statistical uncertainty
16
uncertainty
9
monte carlo
8
accumulated doses
8
prostate cancer
8
modes variation
8
translations local-rigid
8
local-rigid prostate
8
uncertainty setting
8
reduced calculation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!