Objectives: Transcutaneous carbon dioxide (Tc co2 ) monitoring can noninvasively assess ventilation by estimating carbon dioxide ( CO2 ) levels in the blood. We aimed to evaluate the accuracy of Tc co2 monitoring in critically ill children by comparing it to the partial pressure of arterial carbon dioxide (Pa co2 ). In addition, we sought to determine the variation between Tc co2 and Pa co2 acceptable to clinicians to modify patient care and to determine which patient-level factors may affect the accuracy of Tc co2 measurements.

Design: Retrospective observational cohort study.

Setting: Single, quaternary care PICU from July 1, 2012, to August 1, 2020.

Patients: Included participants were admitted to the PICU and received noninvasive ventilation support (i.e., continuous or bilevel positive airway pressure), conventional mechanical ventilation, or high-frequency oscillatory or percussive ventilation with Tc co2 measurements obtained within 15 minutes of Pa co2 measurement.

Interventions: None.

Measurements And Main Results: Three thousand four hundred seven paired arterial blood gas and Tc co2 measurements were obtained from 264 patients. Bland-Altman analysis revealed a bias of -4.4 mm Hg (95% CI, -27 to 18.3 mm Hg) for Tc co2 levels against Pa co2 levels on the first measurement pair for each patient, which fell within the acceptable range of ±5 mm Hg stated by surveyed clinicians, albeit with wide limits of agreement. The sensitivity and specificity of Tc co2 to diagnose hypercarbia were 93% and 71%, respectively. Vasoactive-Infusion Score (VIS), age, and self-identified Black/African American race confounded the relationship between Tc co2 with Pa co2 but percent fluid overload, weight-for-age, probe location, and severity of illness were not significantly associated with Tc co2 accuracy.

Conclusions: Tc co2 monitoring may be a useful adjunct to monitor ventilation in children with respiratory failure, but providers must be aware of the limitations to its accuracy.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11368163PMC
http://dx.doi.org/10.1097/PCC.0000000000003564DOI Listing

Publication Analysis

Top Keywords

co2
17
carbon dioxide
16
dioxide co2
12
co2 monitoring
12
co2 levels
12
transcutaneous carbon
8
monitoring critically
8
critically ill
8
ill children
8
accuracy co2
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!