Background: The first step in computed tomography (CT) reconstruction is to estimate attenuation pathlength. Usually, this is done with a logarithm transformation, which is the direct solution to the Beer-Lambert Law. At low signals, however, the logarithm estimator is biased. Bias arises both from the curvature of the logarithm and from the possibility of detecting zero counts, so a data substitution strategy may be employed to avoid the singularity of the logarithm. Recent progress has been made by Li et al. [IEEE Trans Med Img 42:6, 2023] to modify the logarithm estimator to eliminate curvature bias, but the optimal strategy for mitigating bias from the singularity remains unknown.

Purpose: The purpose of this study was to use numerical techniques to construct unbiased attenuation pathlength estimators that are alternatives to the logarithm estimator, and to study the uniqueness and optimality of possible solutions, assuming a photon counting detector.

Methods: Formally, an attenuation pathlength estimator is a mapping from integer detector counts to real pathlength values. We constrain our focus to only the small signal inputs that are problematic for the logarithm estimator, which we define as inputs of <100 counts, and we consider estimators that use only a single input and that are not informed by adjacent measurements (e.g., adaptive smoothing). The set of all possible pathlength estimators can then be represented as points in a 100-dimensional vector space. Within this vector space, we use optimization to select the estimator that (1) minimizes mean squared error and (2) is unbiased. We define "unbiased" as satisfying the numerical condition that the maximum bias be less than 0.001 across a continuum of 1000 object thicknesses that span the desired operating range. Because the objective function is convex and the constraints are affine, optimization is tractable and guaranteed to converge to the global minimum. We further examine the nullspace of the constraint matrix to understand the uniqueness of possible solutions, and we compare the results to the Cramér-Rao bound of the variance.

Results: We first show that an unbiased attenuation pathlength estimator does not exist if very low mean detector signals (equivalently, very thick objects) are permitted. It is necessary to select a minimum mean detector signal for which unbiased behavior is desired. If we select two counts, the optimal estimator is similar to Li's estimator. If we select one count, the optimal estimator becomes non-monotonic. The oscillations cause the unbiased estimator to be noise amplifying. The nullspace of the constraint matrix is high-dimensional, so that unbiased solutions are not unique. The Cramér-Rao bound of the variance matches well with the expected scaling law and cannot be attained.

Conclusion: If arbitrarily thick objects are permitted, an unbiased attenuation pathlength estimator does not exist. If the maximum thickness is restricted, an unbiased estimator exists but is not unique. An optimal estimator can be selected that minimizes variance, but a bias-variance tradeoff exists where a larger domain of unbiased behavior requires increased variance.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11659049PMC
http://dx.doi.org/10.1002/mp.17406DOI Listing

Publication Analysis

Top Keywords

attenuation pathlength
16
logarithm estimator
16
unbiased attenuation
8
pathlength estimators
8
photon counting
8
logarithm
7
pathlength
5
estimator
5
existence uniqueness
4
uniqueness efficiency
4

Similar Publications

Background: The first step in computed tomography (CT) reconstruction is to estimate attenuation pathlength. Usually, this is done with a logarithm transformation, which is the direct solution to the Beer-Lambert Law. At low signals, however, the logarithm estimator is biased.

View Article and Find Full Text PDF

Significance: Our goal is to understand the root cause of reported oxygen saturation ( ) overestimation in heavily pigmented skin types to devise solutions toward enabling equity in pulse oximeter designs.

Aim: We aim to gain theoretical insights into the effect of skin tone on curves using a three-dimensional, four-layer tissue model representing a finger.

Approach: A finger tissue model, comprising the epidermis, dermis, two arteries, and a bone, was developed using a Monte Carlo-based approach in the MCmatlab software.

View Article and Find Full Text PDF

Detecting Contaminants in Water Based on Full Scattering Profiles within the Single Scattering Regime.

ACS Omega

July 2023

The Faculty of Engineering and the Institute of Nanotechnology and Advanced Materials, Bar Ilan University, Ramat Gan 5290000, Israel.

Clean water is essential for maintaining human health. To ensure clean water, it is important to use sensitive detection methods that can identify contaminants in real time. Most techniques do not rely on optical properties and require calibrating the system for each level of contamination.

View Article and Find Full Text PDF

Continuous measurement of the concentrations of ethanol and its metabolites, acetaldehyde and acetic acid, is important for the alcohol research community. Most studies only measure ethanol because accurate measurement of all three compounds is challenging. Measurement inside tissue/brain is done using a microdialysis technique, followed by off-line analysis using gas chromatography (GC).

View Article and Find Full Text PDF

Mid-IR dispersion spectroscopy - A new avenue for liquid phase analysis.

Spectrochim Acta A Mol Biomol Spectrosc

February 2023

Research Division of Environmental Analytics, Process Analytics and Sensors, Institute of Chemical Technologies and Analytics, Technische Universität Wien, Getreidemarkt 9/164-UPA, 1060 Vienna, Austria. Electronic address:

Mid-IR dispersion spectroscopy is an attractive, novel approach to liquid phase analysis that extends the possibilities of traditional methods based on the detection of absorption via intensity attenuation. This technique detects inherent refractive index changes (phase shifts) induced by IR light interaction with absorbing matter. In contrast to classic absorption spectroscopy, it provides extended dynamic range, baseline-free detection, constant sensitivity, and inherent immunity to power fluctuation.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!