The scattered-light time-history diagnostic (SLTD) suite measures time-resolved scattered light in three wavelength bands: stimulated Brillouin scattering (350-352 nm), stimulated Raman scattering (430-760 nm), and plasma emission at half the laser frequency (695-735 nm), at 15 locations around the National Ignition Facility (NIF) target chamber. The SLTD, along with the full-aperture backscatter station (FABS), collects scattered light from direct- and indirect-drive inertial confinement fusion experiments. The SLTD calibration was revisited after a discrepancy between FABS and SLTD measurements was observed on NIF polar direct-drive [Skupsky et al., Phys. Plasmas 11, 2763 (2004)] experiments. An integrated calibration of the SLTD was performed for the first time, and individual components were also calibrated for the wavelengths of 351, 527, and 532 nm. The optical transmission of the instrument was measured to be (1.12 ± 0.04) × 10-7 and (1.96 ± 0.11) × 10-7 for the wavelengths of 351 and 532 nm, respectively. The revised calibration at 351 nm brings the SLTD measured scattered energy in agreement with the FABS measured scattered energy after additionally accounting for the degradation of an optical element in FABS. This decreased the inferred absorption by 7% for a representative experiment. However, discrepancies remain between FABS and SLTD measurements in the SRS band (532 nm).
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1063/5.0218987 | DOI Listing |
Rev Sci Instrum
September 2024
Laboratory for Laser Energetics, University of Rochester, Rochester, New York 14623-1299, USA.
The scattered-light time-history diagnostic (SLTD) suite measures time-resolved scattered light in three wavelength bands: stimulated Brillouin scattering (350-352 nm), stimulated Raman scattering (430-760 nm), and plasma emission at half the laser frequency (695-735 nm), at 15 locations around the National Ignition Facility (NIF) target chamber. The SLTD, along with the full-aperture backscatter station (FABS), collects scattered light from direct- and indirect-drive inertial confinement fusion experiments. The SLTD calibration was revisited after a discrepancy between FABS and SLTD measurements was observed on NIF polar direct-drive [Skupsky et al.
View Article and Find Full Text PDFRev Sci Instrum
March 2021
Laboratory for Laser Energetics, University of Rochester, Rochester, New York 14623, USA.
The Scattered Light Time-history Diagnostic (SLTD) is being implemented at the National Ignition Facility (NIF) to greatly expand the angular coverage of absolute scattered-light measurements for direct- and indirect-drive inertial confinement fusion (ICF) experiments. The SLTD array will ultimately consist of 15 units mounted at a variety of polar and azimuthal angles on the NIF target chamber, complementing the existing NIF backscatter suite. Each SLTD unit collects and diffuses scattered light onto a set of three optical fibers, which transport the light to filtered photodiodes to measure scattered light in different wavelength bands: stimulated Brillouin scattering (350 nm-352 nm), stimulated Raman scattering (430 nm-760 nm), and ω/2 (695 nm-745 nm).
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!