Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: The VASCADE MVP venous vascular closure system is commonly used for percutaneous venotomy closure in catheter-based procedures utilizing sheath sizes 6-12 French. However, its application with larger sheaths such as ones required in left atrial appendage occlusion (LAAO) has yet to be explored.
Aims: This study compared the efficacy and safety of VASCADE MVP versus conventional Figure-of-8 sutures (Fo8) for femoral venotomy closure in patients undergoing Watchman LAAO.
Methods: This single-center retrospective analysis included patients who underwent post-LAAO femoral venotomy closure with either VASCADE MVP or Fo8 sutures. The primary efficacy endpoint was time to hemostasis (TTH). Primary safety endpoints were 30-day access site-related readmissions and major complications. Minor access-site-related complications were also assessed.
Results: 107 patients underwent post-LAAO femoral venotomy closure with VASCADE MVP, of which 101 were successful (94.4%). 99 patients underwent conventional closure with Fo8 sutures and supplemental manual pressure. Baseline characteristics were similar between groups. TTH was significantly shorter with VASCADE MVP (152.9 vs. 335.8 s, p = 0.001). Major safety outcomes were comparable. However, the conventional group experienced a higher incidence of ecchymosis (16.2% vs 4.7%, p = 0.007) and 30-day all-cause readmission (18.2% vs 8.4%, p = 0.038).
Conclusions: VASCADE MVP significantly reduced TTH compared to Fo8 sutures after LAAO. While demonstrating a similar safety profile, VASCADE MVP may offer a more efficient approach to femoral vein closure in this patient population.