Low-cost, portable devices capable of accurate physiological measurements are attractive tools for coaches, athletes, and practitioners. The purpose of this study was primarily to establish the validity and reliability of Movesense HR+ ECG measurements compared to the criterion three-lead ECG, and secondarily, to test the industry leader Garmin HRM. Twenty-one healthy adults participated in running and cycling incremental test protocols to exhaustion, both with rest before and after. Movesense HR+ demonstrated consistent and accurate R-peak detection, with an overall sensitivity of 99.7% and precision of 99.6% compared to the criterion; Garmin HRM sensitivity and precision were 84.7% and 87.7%, respectively. Bland-Altman analysis compared to the criterion indicated mean differences (SD) in RR' intervals of 0.23 (22.3) ms for Movesense HR+ at rest and 0.38 (18.7) ms during the incremental test. The mean difference for Garmin HRM-Pro at rest was -8.5 (111.5) ms and 27.7 (128.7) ms for the incremental test. The incremental test correlation was very strong ( = 0.98) between Movesense HR+ and criterion, and moderate ( = 0.66) for Garmin HRM-Pro. This study developed a robust peak detection algorithm and data collection protocol for Movesense HR+ and established its validity and reliability for ECG measurement.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11397956PMC
http://dx.doi.org/10.3390/s24175713DOI Listing

Publication Analysis

Top Keywords

movesense hr+
24
incremental test
16
validity reliability
12
compared criterion
12
reliability movesense
8
hr+ ecg
8
ecg measurements
8
running cycling
8
garmin hrm
8
garmin hrm-pro
8

Similar Publications

Validity and Reliability of Movesense HR+ ECG Measurements for High-Intensity Running and Cycling.

Sensors (Basel)

September 2024

School of Physical Education, Sport and Exercise Sciences, University of Otago, Dunedin 9054, New Zealand.

Low-cost, portable devices capable of accurate physiological measurements are attractive tools for coaches, athletes, and practitioners. The purpose of this study was primarily to establish the validity and reliability of Movesense HR+ ECG measurements compared to the criterion three-lead ECG, and secondarily, to test the industry leader Garmin HRM. Twenty-one healthy adults participated in running and cycling incremental test protocols to exhaustion, both with rest before and after.

View Article and Find Full Text PDF

Sensorized T-Shirt with Intarsia-Knitted Conductive Textile Integrated Interconnections: Performance Assessment of Cardiac Measurements during Daily Living Activities.

Sensors (Basel)

November 2023

Textile Materials Technology, Department of Textile Technology, Faculty of Textiles, Engineering and Business Swedish School of Textiles, University of Borås, 503 32 Borås, Sweden.

The development of smart wearable solutions for monitoring daily life health status is increasingly popular, with chest straps and wristbands being predominant. This study introduces a novel sensorized T-shirt design with textile electrodes connected via a knitting technique to a Movesense device. We aimed to investigate the impact of stationary and movement actions on electrocardiography (ECG) and heart rate (HR) measurements using our sensorized T-shirt.

View Article and Find Full Text PDF

Estimation of Respiratory Frequency in Women and Men by Kubios HRV Software Using the Polar H10 or Movesense Medical ECG Sensor during an Exercise Ramp.

Sensors (Basel)

September 2022

Interdisciplinary Institute of Exercise Science and Sports Medicine, MSH Medical School Hamburg, University of Applied Sciences and Medical University, Am Kaiserkai 1, 20457 Hamburg, Germany.

Monitoring of the physiologic metric, respiratory frequency (RF), has been shown to be of value in health, disease, and exercise science. Both heart rate (HR) and variability (HRV), as represented by variation in RR interval timing, as well as analysis of ECG waveform variability, have shown potential in its measurement. Validation of RF accuracy using newer consumer hardware and software applications have been sparse.

View Article and Find Full Text PDF

The value of heart rate variability (HRV) in the fields of health, disease, and exercise science has been established through numerous investigations. The typical mobile-based HRV device simply records interbeat intervals, without differentiation between noise or arrythmia as can be done with an electrocardiogram (ECG). The intent of this report is to validate a new single channel ECG device, the Movesense Medical sensor, against a conventional 12 channel ECG.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!