Background: This study aimed to compare Doppler (the most commonly used technique for non-invasive indirect blood pressure [NIBP] measurement in rabbits) and oscillometric devices for NIBP measurements in conscious rabbits.

Methods: Blood pressure (BP) measurements were obtained for 49 conscious rabbits using the Doppler and oscillometric devices. Each measurement was repeated in triplicate. Three rabbits were excluded from the study because it was not possible to obtain BP measurements using the oscillometric device. The American College of Veterinary Internal Medicine (ACVIM) guidelines were followed to compare the results obtained with the two devices.

Results: A total of 164 systolic BP measurements were obtained with the Doppler device and 182 were obtained with the oscillometric technique. The mean of the three values obtained for each animal with each device was used for statistical analysis. The analysis demonstrated a poor level of agreement between the measurements obtained with the two devices.

Limitations: Oscillometric devices rely on built-in software algorithms to perform calculations and produce their measurements. However, the oscillometric devices currently available only have the software for dogs and cats installed. Therefore, these conclusions should be revised when a device with software specifically for rabbits is commercialised.

Conclusions: This study indicates that oscillometric devices should not be used as a substitute for the Doppler technique when measuring BP in rabbits.

Download full-text PDF

Source
http://dx.doi.org/10.1002/vetr.4399DOI Listing

Publication Analysis

Top Keywords

oscillometric devices
24
blood pressure
12
conscious rabbits
8
oscillometric
8
measurements oscillometric
8
rabbits
6
devices
6
measurements
6
doppler
5
non-invasive blood
4

Similar Publications

Evaluation of Arterial Stiffness Parameters Measurement With Noninvasive Methods-A Systematic Review.

Cardiol Res Pract

December 2024

Department of Family Medicine, Medical University of Białystok, Podlaskie Voivodeship, 15-054 Białystok, Poland.

Arterial stiffness, as determined by pulse wave velocity (PWV), is a recognized marker of cardiovascular risk. Noninvasive technologies have enabled easier and more accessible assessments of PWV. The current gold standard for measuring carotid-femoral PWV (cfPWV)-a reliable indicator of arterial stiffness-utilizes applanation tonometry devices, as recommended by the Artery Society Guidelines.

View Article and Find Full Text PDF

To (1) assess the accuracy of the Uscom BP+ oscillometric upper-arm professional blood pressure (BP) monitor in children and adolescents as per the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018); (2) assess the suitability of a non-standard cuff; and (3) assess the impact of motion artefact on device accuracy and precision. Children and adolescents were recruited to fulfil the sex/cuff distribution criteria of the Universal Standard. Single arm sequential BP measurements were used, the test device measured on deflate with an altered child/adolescent algorithm, using three standard (Extra Small, Small Adult, Adult) and, as exploratory analysis, additionally in one non-standard cuff (Wide Range; WR) (arm circumferences: 12-17, 16-24, 22-32, 22-42 cm, respectively).

View Article and Find Full Text PDF

Background: The systolic blood pressure/workload (SBP/MET) slope was recently reported to reliably identify an exaggerated blood pressure response (eBPR) in the normal population and in athletes. We investigated whether the aortic root diameter (AoD) also correlates with an eBPR and vascular function in elite athletes.

Methods: We examined 652 healthy male elite athletes (age 25.

View Article and Find Full Text PDF

Background: Oscillometry devices allow quantification of respiratory function at tidal breathing but device-specific reference equations are scarce: the present study aims to create sex-specific oscillometric reference values for children and adolescents using the Resmon PRO FULL device.

Methods: Healthy participants (n=981) aged 6 to 17 years of the Austrian LEAD general population cohort were included. Subjects had normal weight (body mass index ≤99th percentile) and normal lung volumes (total lung capacity (TLC) ≥ lower limit of normal).

View Article and Find Full Text PDF

Association of different domains of sedentary behavior and cardiovascular risk factors in adolescents: Cross-sectional study.

Hipertens Riesgo Vasc

November 2024

School of Technology and Sciences, São Paulo State University (Unesp), Presidente Prudente, São Paulo, Brazil. Electronic address:

Background: Sedentary behavior (SB) has been related to cardiovascular risk factors (CVRF) such as high BMI, waist circumference (WC) and blood pressure (BP), including pediatric populations. However, it is still unclear whether the association between SB and CVRF could be domain dependent. Therefore, this study aimed to analyze the relationship between sedentary at different domains (time spent in TV, videogames, computer, smartphone) with CVRF in adolescents.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!