Statement Of Problem: The biomechanical stability of the implant-prosthesis assembly and its maintenance under function is a determining factor in the success of implant prosthesis rehabilitation, but studies of different angular tapers are lacking.

Purpose: The purpose of this in vitro study was to evaluate the axial displacement and torque loss of prosthetic abutments with Morse-type connections of different angular tapers after thermomechanical cycling by using microcomputed tomography (µCT) and a digital torque wrench.

Material And Methods: Eighteen Ø3.5×11.5-mm implants were embedded in polyvinyl chloride cylinders, and the 3 different types of abutments (n=6) with angular tapers of 11.5 degrees - Alvim Cone Morse (ACM-11.5), 16.0 degrees - Helix Gran Morse (HGM-16), and 24.0 degrees - Nobel Replace Conical Connection (NRC-24) were installed with the torque recommended by the manufacturers. To calculate the axial displacements of the abutments about the implants, zirconia maxillary canine crowns were fabricated using a computer-aided design and computer-aided manufacturing (CAD-CAM) system and cemented onto the abutments. Before and after thermomechanical cycling, the assemblies were scanned using microtomography (micro-CT) to assess axial displacement, and the torque losses were calculated after the abutments were finally unscrewed. A load of 100 N, frequency of 2 Hz, and 10 cycles with temperature variation of 5 °C to 55 °C were used for the thermomechanical cycling. Analysis of variance and the Tukey post hoc test (α=.05) were used for analysis.

Results: A significant difference was observed between the abutments for axial displacement, measured after thermomechanical cycling (P=.002). The ACM-11.5 abutment showed the highest mean value (134.1 ±58.7 µm), different from HGM-16 (63.3 ±26.1 µm) (P=.013) and NRC-24 (42.7 ±8.7 µm) (P=.002); the 2 latter groups were similar to each other (P=.618). For the torque losses, no significant difference was found among the abutments (P=.928), but there were significant differences for the thermomechanical cycling (before and after) (P<.001) in that the loss of torque was greater after thermocycling.

Conclusions: The smaller the taper angle of the Morse-type prosthetic abutment, the greater its axial displacement, and the thermomechanical cycling significantly reduced pretorque, regardless of taper.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.prosdent.2024.08.006DOI Listing

Publication Analysis

Top Keywords

thermomechanical cycling
20
axial displacement
16
angular tapers
16
displacement torque
12
torque loss
8
abutments
8
prosthetic abutments
8
torque losses
8
torque
6
thermomechanical
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!