  • The study evaluated a redesigned medial meniscus prosthesis and its fixation technique after issues were noted in a prior trial.
  • Ten patients with postmeniscectomy pain received the new prosthesis, and their outcomes were tracked over 12 months using patient-reported measures and imaging.
  • While the redesign showed some clinical improvements, four prostheses failed, indicating that further work on positioning and fixation methods is needed to enhance safety and effectiveness.

Article Abstract

Purpose: After unsatisfactory results in a first-in-human clinical investigation with an anatomically shaped medial meniscus prosthesis, the prosthesis and its fixation technique were altered. This interim analysis of a prospective single-arm clinical investigation aims to evaluate safety and clinical performance in a first-in-human study with the redesigned meniscus prosthesis system.

Methods: Ten patients suffering from medial postmeniscectomy pain syndrome were treated with the meniscus prosthesis. Patient-reported outcome measures were obtained at baseline and at 6-week, 3-, 6-, and 12-month follow-up. Radiographs and magnetic resonance imaging scans were obtained to evaluate joint degeneration and prosthesis location.

Results: The device alterations resolved the issues occurring with the previous design, but four prostheses were explanted after fixation failure or subluxation of the prosthesis. Five out of six patients who reached 1-year follow-up reported a clinically significant improvement of the knee injury and osteoarthritis outcome score pain subscale. Imaging revealed no adverse effects on joint degeneration.

Conclusion: The failure mechanisms that occurred with the previous design have been resolved, but the new fixation technique introduced new safety issues. Improvement of positioning and fixation techniques are considered essential future adaptations to reduce the risk of failure. The good clinical outcomes reported by the patients reaching 1-year follow-up indicate that the medial meniscus prosthesis is a potential solution for patients suffering from postmeniscectomy pain syndrome.

Level Of Evidence: Level II.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ksa.12454DOI Listing

Publication Analysis

Top Keywords

meniscus prosthesis
20
medial meniscus
12
1-year follow-up
12
clinical performance
8
prosthesis
8
clinical investigation
8
fixation technique
8
patients suffering
8
postmeniscectomy pain
8
previous design
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!