Introduction and Validation of Low-Cost Ocular Biometry in Healthy and Cataractous Eyes Using a Thermally Tunable Swept-Source Laser.

Am J Ophthalmol

Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria; Christian Doppler Laboratory for Innovative Optical Imaging and its Translation to Medicine, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria. Electronic address:

Published: January 2025

AI Article Synopsis

Article Abstract

Purpose: To introduce and validate a novel substantially lower-priced and rapid swept-source investigational optical biometer in healthy and cataractous eyes, using a thermally tuned laser diode used extensively in cell phones and data communication as an alternative swept source.

Design: Prospective accuracy, validity, and reliability analysis.

Methods: A total of 60 eyes of 59 subjects (29 eyes of 29 healthy subjects and 31 eyes of 30 cataract patients) were enrolled in a prospective comparative study at the Vienna General Hospital between August 2021 and April 2023. Averaged intraocular distances were acquired in 2.5 seconds from datasets consisting of 5000 consecutive A-scans at a single position by a low-cost swept-source optical biometry (SSOB) system. Instrument repeatability was assessed via standard deviations (SDs) and coefficients of variation (CoVs) of parameters such as axial length (AL), anterior chamber depth (ACD), lens thickness (LT), and central corneal thickness (CCT). Healthy subjects and cataract patients were subsequently measured on the same day with the SSOB and a referential partial coherence interferometry (PCI) biometer (IOLMaster 500) to establish AL inter-device correlation (r) for instrument calibration. AL and ACD as shared parameters between both biometers were evaluated for their limits of agreements (LoA) using Bland-Altman analysis.

Results: Repeated measurements of AL, ACD, LT, and CCT revealed SDs of 18 µm, 12 µm, 12 µm, and 10 µm, respectively. All parameters except for CCT had a COV <1%. Except for 1 eye with white cataract, 59 eyes of 59 study participants with various degrees and types of cataract could be measured with both devices. The AL inter-device correlation was excellent (r > 0.99). The 95% LoAs between both biometers were -0.14 to 0.13 mm for AL and -0.28 to 0.25 mm for ACD.

Conclusions: Optical biometry using a thermally tunable VCSEL swept-source light source has the potential to provide clinically relevant biometric parameters at an unprecedented 100-fold lower price point than currently used state-of-the-art optical biometers, paving the way for compact devices in remote care settings.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajo.2024.08.026DOI Listing

Publication Analysis

Top Keywords

µm µm
12
healthy cataractous
8
cataractous eyes
8
eyes thermally
8
thermally tunable
8
subjects eyes
8
healthy subjects
8
cataract patients
8
optical biometry
8
eyes
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!