A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Analytical verification of the Atellica VTLi point of care high sensitivity troponin I assay. | LitMetric

Objectives: The Siemens Point-of-Care Testing (POC) Atellica VTLi high-sensitivity troponin I (hsTnI) device has been previously validated. Verification independently provides evidence that an analytical procedure fulfils concordance with laboratory assays, imprecision, and hemolysis interference requirements.

Methods: Five whole blood samples spanning the measuring interval were analysed 20 times in succession. Hemolysis interference was assessed at three troponin concentrations by spiking five hemolysate concentrations to plasma to achieve free hemoglobin concentrations 35-1,000 mg/dL. Concordance between whole blood (VTLi) and plasma on laboratory analysers (Beckman, Roche, Siemens) was assessed by Pearson correlation and kappa statistics at the (LOQ) and upper reference limit (URL). This was repeated for frozen plasma samples.

Results: Coefficients of variation for whole blood were <10 % for whole blood troponin concentrations of 9.2 and 15.9 ng/L, thus below the URL. Hemolysis positively interfered; at 250 mg/dL affecting the low troponin sample (+3 ng/L; +60 %) and high troponin sample (+37 ng/L; +24 %). Correlation coefficients were 0.98, 0.90 and 0.97 between VTLi and Beckman, Roche and Siemens assays respectively. Corresponding kappa statistics were 0.80, 0.73 and 0.84 at the LOQ and 0.70, 0.44 and 0.67 at the URL.

Conclusions: Concordances between VTLi and laboratory assays were at least non-inferior to those between laboratory assays. Imprecision met manufacturer claims and was consistent with a high sensitivity assay. There is potential for hemolysis interference, highlighting the need for quality samples. The results support performance characteristics previously reported in validation studies, and the device offers acceptable performance for use within intended medical settings.

Download full-text PDF

Source
http://dx.doi.org/10.1515/cclm-2024-0312DOI Listing

Publication Analysis

Top Keywords

atellica vtli
8
hemolysis interference
8
analytical verification
4
verification atellica
4
vtli point
4
point care
4
care high
4
high sensitivity
4
sensitivity troponin
4
troponin assay
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!