Purpose: Determination of the appropriate ablative parameters is the key to the success and safety of microwave ablation (MWA) of lung tumors. The purpose of this study was to provide guidelines and recommendations for the optimal time and power for lung tumor MWA.
Material And Methods: MWA using a 2450-MHz system was evaluated in a porcine lung. The independent variables were power (30, 40, 50, 60, 70, and 80 W) and time (2, 4, 6, 8, 10, and 12 min), and the outcome variable was the volume of ablation. Lung tissues were procured after MWA for measurement and histological evaluation. Analysis of variance was used for statistical analysis, followed by least significant difference (LSD) t-tests where appropriate. A P value of <0.05 was considered statistically significant.
Results: The outcome variable (ablative volume) was significantly affected by time, power, and time/power interaction (P < 0.05). When the total output energy was kept constant, the combination of higher power and shorter time obtained a larger ablative volume, especially in the low- and medium-energy groups (P < 0.01).
Conclusions: We propose guidelines for ablative volume based on different time and power variables to provide a reference for clinical applications.