Background: MCA bifurcation aneurysms pose treatment challenges because of the complex hemodynamics at the bifurcation and the risk of rupture. FDS implantation has been controversial and there are only limited reports. Therefore, the aim of this study was to assess the efficacy and safety of this treatment strategy using p64 MW HPC and p48 MW HPC FDSs for MCA bifurcation aneurysms, compared with the p64 classic FDS.

Materials And Methods: We retrospectively analyzed our institutional database and identified all patients with saccular, non-ruptured MCA bifurcation aneurysms treated with p64 MW HPC, p48 MW HPC, or p64 classic FDS implantation alone. Aneurysms with implantation of additional devices in the same session, previous treatments, and acutely ruptured and fusiform aneurysms were excluded.

Results: A total of 79 aneurysms met the inclusion criteria: 23 receiving a p64 MW HPC, 34 receiving a p48 MW HPC, and 22 receiving a p64 classic FDS. The occlusion rate was highest for the p48 MW HPC 2 mm FDS, at 88.9% at FU2, compared with 72.2% for the p64 MW HPC and 70.6% for the p64 classic. The time to aneurysm occlusion was shortest with the p64 MW HPC, at 178.31 days. The highest retreatment rate was observed with the p48 MW HPC 3 mm.

Conclusion: Treatment of MCA bifurcation aneurysms with a p48 MW HPC 2 mm or p64 MW HPC FDS is a safe and reliable strategy achieving high aneurysm occlusion rates - attributable to their lower porosity in relation to the parent vessel diameter as compared to the p48 MW HPC 3 mm FDS-, with reasonable morbidity and mortality.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11349692PMC
http://dx.doi.org/10.3389/fneur.2024.1415861DOI Listing

Publication Analysis

Top Keywords

p48 hpc
32
p64 hpc
28
mca bifurcation
20
bifurcation aneurysms
20
p64 classic
20
hpc
15
p64
12
hpc p48
12
aneurysms
8
p48
8

Similar Publications

Background: MCA bifurcation aneurysms pose treatment challenges because of the complex hemodynamics at the bifurcation and the risk of rupture. FDS implantation has been controversial and there are only limited reports. Therefore, the aim of this study was to assess the efficacy and safety of this treatment strategy using p64 MW HPC and p48 MW HPC FDSs for MCA bifurcation aneurysms, compared with the p64 classic FDS.

View Article and Find Full Text PDF

Background: This study reports a multicenter experience of using hydrophilic polymer-coated (HPC) flow diverters with prasugrel single antiplatelet therapy to treat ruptured aneurysms with subarachnoid hemorrhage (SAH).

Methods: Patients treated for intracranial aneurysms within 30 days after SAH with a p64/p48 MW HPC flow diverter were prospectively identified. Clinical presentation and outcomes, periprocedural and postprocedural complications, and degree of occlusion at follow-up were evaluated.

View Article and Find Full Text PDF

Glanzmann Thrombasthenia (GT) is an inherited platelet disorder caused by defects in platelet integrin αβ (GPIIb/IIIa), which is a platelet receptor essential for the binding of fibrinogen. This can lead to severe bleeding, especially after trauma or perioperatively, and to microcytic anemia because of chronic blood loss. We report on a 40-year-old female patient with extensive bleeding complications and platelet antibody formation who presented in Homburg and Freiburg for extensive platelet function analyses and molecular genetic analyses.

View Article and Find Full Text PDF

Flow diverter (FD) is increasingly used in the management of wide necked cerebral aneurysms. Despite a reported lower efficacy in middle cerebral artery (MCA) aneurysms, they are still being utilised. Microsurgery is best considered as an index treatment, but can also be a safe and effective treatment when encountering a persistent MCA aneurysm after prior FD.

View Article and Find Full Text PDF
Article Synopsis
  • Flow diversion (FD) is a key treatment for large, wide-necked aneurysms but requires dual antiplatelet therapy (DAPT) to manage thromboembolic risks, complicating patient care due to potential hemorrhagic and nuisance complications.
  • * Newer FD technologies incorporate surface modifications for better blood compatibility, aiming to reduce clotting and potentially allow the use of a single antiplatelet drug.
  • * This review addresses the molecular mechanisms and clinical data on four major flow diverters with advanced coatings to provide a comprehensive reference for neurointerventional aneurysm treatment, excluding the uncoated Balt silk device.
View Article and Find Full Text PDF