Mercury (Hg) is a toxic element which impacts on biological systems and ecosystems. Because the toxicity of Hg species is highly dependent on their concentration levels and chemical forms, the sensitive identification of the chemical forms of Hg-i.e., Hg speciation-is of major significance in providing meaningful information about the sources of Hg exposure. In this study, a microfluidic-based device made of high-clarity poly(methyl methacrylate) (PMMA) was fabricated. Then, titanium dioxide nanoparticles (nano-TiOs) were attached to the treated channel's interior with the aid of poly(diallyldimethylammonium chloride) (PDADMAC). After coupling the nano-TiO-coated microfluidic-based photocatalyst-assisted reduction device (the nano-TiO-coated microfluidic-based PCARD) with high-performance liquid chromatography (HPLC) and inductively coupled plasma mass spectrometry (ICP-MS), a selective and sensitive, hyphenated system for Hg speciation was established. Validation procedures demonstrated that the method could be satisfactorily applied to the determination of mercury ions (Hg) and methylmercury ions (CHHg) in both human urine and water samples. Remarkably, the zeta potential measured clearly indicated that the PDADMAC-capped nano-TiOs with a predominance of positive charges indeed provided a steady force for firm attachment to the negatively charged device channel. The cause of the durability of the nano-TiO-coated microfluidic-based PCARD was clarified thus.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11360531 | PMC |
http://dx.doi.org/10.3390/polym16162366 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!