Exercise-induced muscle injury is one of the most common types of sports injuries. Skeletal muscle troponin I (skTnI) serves as an ideal biomarker in assessing such injuries, facilitating timely detection and evaluation. In this study, we develop a fluorescent sandwich lateral flow immunoassay (LFIA) combined with a desktop analyzer for rapid detection of skTnI. Through optimizing the reaction system, the assay achieves a satisfying detection performance, reaching a limit of detection (LOD) of 0.5 ng/mL with a turnaround time of 15 min. The proposed detection platform offers portability, ease of use, and high sensitivity, which facilitates the monitoring of exercise-induced muscle injuries at the point of care. This feature is particularly advantageous for end users, enabling timely detection of sports-related injuries and ultimately enhancing prognosis and sports life.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11352793PMC
http://dx.doi.org/10.3390/bios14080381DOI Listing

Publication Analysis

Top Keywords

lateral flow
8
flow immunoassay
8
skeletal muscle
8
muscle troponin
8
muscle injury
8
point care
8
exercise-induced muscle
8
timely detection
8
detection
7
muscle
5

Similar Publications