AI Article Synopsis

  • * Twenty patients were enrolled, and results after 8 months showed minimal in-stent late lumen loss (0.7 mm) and complete stent strut coverage, indicating successful healing.
  • * There were no adverse cardiovascular events reported within the same period, suggesting that the new stent design is safe and effective in treating coronary artery disease.

Article Abstract

Background: In patients with coronary artery disease treated with permanent polymer-coated drug-eluting stents (DES), the persistent presence of a less biocompatible polymer might delay arterial healing. Thin strut polymer-free DES have the potential to improve clinical outcomes and reduce the duration of dual antiplatelet therapy (DAPT). The purpose of this first-in-human study was to assess the safety and effectiveness of a novel polymer-free DES in patients with de novo coronary lesions. The TIGERevolutioN® stent (CG Bio Co., Ltd., Seoul, Korea) consists of a cobalt chromium platform with a strut thickness of 70 μm and a surface treated with titanium dioxide onto which everolimus-eluting stent (EES) is applied abluminally (6 µg/mm of stent length) without utilization of a polymer.

Methods: A total of 20 patients were enrolled, with de novo coronary lesions (stable or unstable angina) and > 50% diameter stenosis in a vessel 2.25 to 4.00 mm in diameter and ≤ 40 mm in length for angiographic, optical coherence tomography (OCT), and clinical assessment at 8 months. All patients received DAPT after stent implantation. The primary endpoint was angiographic in-stent late lumen loss (LLL) at 8 months.

Results: Twenty patients with 20 lesions were treated with TIGERevolutioN®. At 8 months, in-stent LLL was 0.7 ± 0.4 mm. On OCT, percent area stenosis was 29.2 ± 9.4% and stent strut coverage was complete in all lesions. No adverse cardiovascular event occurred at 8 months.

Conclusion: The new polymer-free EES was safe and effective with low LLL and excellent strut coverage at 8 months of follow-up.

Trial Registration: Trial Registration: Clinical Research Information Service Identifier: KCT0005699.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11347186PMC
http://dx.doi.org/10.3346/jkms.2024.39.e234DOI Listing

Publication Analysis

Top Keywords

everolimus-eluting stent
8
titanium dioxide
8
polymer-free des
8
novo coronary
8
coronary lesions
8
strut coverage
8
stent
6
patients
5
first-in-human evaluation
4
polymer-free
4

Similar Publications

Background: Few data are available on polymer-free drug-eluting stents in patients undergoing percutaneous coronary intervention (PCI).

Aims: We aimed to determine the efficacy and safety of a polymer-free amphilimus-eluting stent (AES), using a reservoir-based technology for drug delivery, compared with a biodegradable-polymer everolimus-eluting stent (EES).

Methods: This was a randomised, investigator-initiated, assessor-blind, non-inferiority trial conducted at 14 hospitals in Italy (ClinicalTrials.

View Article and Find Full Text PDF

Introduction And Objectives: Patients undergoing percutaneous coronary intervention in vessels with moderate-to-severe tortuosity are at higher risk of adverse outcomes, but data are scarce in the era of newer-generation stents. We compared outcomes following percutaneous coronary intervention in vessels with moderate-to-severe tortuosity using a bioresorbable-polymer sirolimus-eluting stent (BP-SES) vs a durable-polymer everolimus-eluting stent.

Methods: A total of 2350 patients from the BIOFLOW II, IV, and V randomized trials were stratified into 2 groups based on target-vessel tortuosity: none-to-mild and moderate-to-severe.

View Article and Find Full Text PDF
Article Synopsis
  • This study investigates the effectiveness of bioresorbable scaffolds (IBS) versus everolimus-eluting stents (EES) for treating below-the-knee arterial disease in dogs, focusing on their support performance and effects on tissue.
  • Both types of implants were successfully used, showing similar levels of arterial narrowing within the first six months, but the IBS scaffold exhibited gradual corrosion over time.
  • Overall, the findings indicate that the IBS is comparable to EES in terms of support and tissue response within 6 months, while demonstrating a manageable corrosion profile over a year.
View Article and Find Full Text PDF

Background: Long-term clinical data on biodegradable-polymer (BP) drug-eluting stents (DES) are limited. The objective of this study was to assess the long-term safety and efficacy of the BP-DES SYNERGY compared to XIENCE V, a durable-polymer (DP)-DES.

Methods: We compared patients treated with BP-DES or DP-DES at our center from 2008 to 2020.

View Article and Find Full Text PDF

Background: Adult Kawasaki patients may require intervention for occlusive coronary artery disease. Some adverse effects of first-generation drug-eluting stent implantation with sirolimus have been reported in this population.

Methods: A total of nine lesions in eight (seven males, one female) patients who underwent stent implantations in this population between 2000 and 2021 were reviewed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!