Objective: Measurement of loop gain in patients with obstructive sleep apnea (OSA) is of great importance for determining the cause of OSA and realizing precision medicine. In this study, automatic measurement of loop gain is carried out during wakefulness based on the pseudorandom binary sequence (PRBS) CO stimulation method.

Methods: A respiratory data acquisition platform is designed and constructed to automate the processes of gas configuration, ventilatory stimulation with CO and data acquisition. The respiratory data are substituted into the model of the ventilatory control system for parameter identification, and the loop gain values are calculated and then compared with the apnea-hypopnea index (AHI) measured in a hypoxia chamber.

Results: A positive correlation is found between loop gain and AHI measured in the hypoxia chamber, with the linear correlation coefficient of approximately 0.65.

Conclusion: The feasibility of automatic measurement of loop gain using the respiratory data acquisition platform based on the PRBS CO stimulation method is validated, and the measured loop gain values can be used to assess the stability of ventilatory control.

Significance: This study provides an automated, rapid, and instrumented solution for loop gain measurement, laying the foundation for wide-scale clinical application of the PRBS CO stimulation method.

Download full-text PDF

Source
http://dx.doi.org/10.1109/TBME.2024.3449410DOI Listing

Publication Analysis

Top Keywords

loop gain
32
measurement loop
16
automatic measurement
12
prbs stimulation
12
respiratory data
12
data acquisition
12
loop
8
gain
8
based pseudorandom
8
pseudorandom binary
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!