Objective: To assess the relationship between intravenous iodinated contrast media (ICM) administration usage and radiation doses for contrast-enhanced (CE) CT of head, chest, and abdomen-pelvis (AP) in international, multicenter settings.

Methods: Our international (n = 16 countries), multicenter (n = 43 sites), and cross-sectional (ConRad) study had two parts. Part 1: Redcap survey with questions on information related to CT and ICM manufacturer/brand and respective protocols. Part 2: Information on 3,258 patients (18-96 years; M:F 1654:1604) who underwent CECT for a routine head (n = 456), chest (n = 528), AP (n = 599), head CT angiography (n = 539), pulmonary embolism (n = 599), and liver CT examinations (n = 537) at 43 sites across five continents. The following information was recorded: hospital name, patient age, gender, body mass index [BMI], clinical indications, scan parameters (number of scan phases, kV), IV-contrast information (concentration, volume, flow rate, and delay), and dose indices (CTDIvol and DLP).

Results: Most routine chest (58.4%) and AP (68.7%) CECT exams were performed with 2-4 scan phases with fixed scan delay (chest 71.4%; AP 79.8%, liver CECT 50.7%) following ICM administration. Most sites did not change kV across different patients and scan phases; most CECT protocols were performed at 120-140 kV (83%, 1979/2685). There were no significant differences between radiation doses for non-contrast (CTDIvol 24 [16-30] mGy; DLP 633 [414-702] mGy·cm) and post-contrast phases (22 [19-27] mGy; 648 [392-694] mGy·cm) (p = 0.142). Sites that used bolus tracking for chest and AP CECT had lower CTDIvol than sites with fixed scan delays (p < 0.001). There was no correlation between BMI and CTDIvol (r ≤ - 0.1 to 0.1, p = 0.931).

Conclusion: Our study demonstrates up to ten-fold variability in ICM injection protocols and radiation doses across different CT protocols. The study emphasizes the need for optimizing CT scanning and contrast protocols to reduce unnecessary contrast and radiation exposure to patients.

Clinical Relevance Statement: The wide variability and lack of standardization of ICM media and radiation doses in CT protocols suggest the need for education and optimization of contrast usage and scan factors for optimizing image quality in CECT.

Key Points: There is a lack of patient-centric CT protocol optimization taking into consideration mainly patients' size. There is a lack of correlation between ICM volume and CT radiation dose across CT protocol. A ten-fold variation in iodine-load for the same CT protocol in sites suggests a lack of standardization.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-024-11017-7DOI Listing

Publication Analysis

Top Keywords

radiation doses
12
scan phases
12
iodinated contrast
8
icm administration
8
fixed scan
8
scan
6
chest
5
sites
5
cect
5
survey radiation
4

Similar Publications

Effect of head size, head material, and radiation dose on the repeatability of CT-RSA measurements of femoral head penetration.

J Mech Behav Biomed Mater

January 2025

Division of Orthopaedic Surgery, Schulich School of Medicine & Dentistry, Western University, London, Ontario, Canada; Department of Medical Biophysics, Schulich School of Medicine & Dentistry, Western University, London, Ontario, Canada. Electronic address:

Background: The risk of early revision of total hip arthroplasty (THA) for polyethylene wear is now low, but there remains a need to perform wear measurements in patients for clinical surveillance. The gold standard of wear measurements has been radiostereometric analysis (RSA), which has limited availability. The use of computed tomography (CT) to perform THA wear measurement was described a decade ago and found to have acceptable accuracy and precision, but high radiation dose was a concern.

View Article and Find Full Text PDF

We report on a procedure for extracting the SPICE model parameters of a RADFET sensor with a dielectric HfO/SiO double-layer. RADFETs, traditionally fabricated as PMOS transistors with SiO, are enhanced by incorporating high-k dielectric materials such as HfO to reduce oxide thickness in modern radiation sensors. The fabrication steps of the sensor are outlined, and model parameters, including the threshold voltage and transconductance, are extracted based on experimental data.

View Article and Find Full Text PDF

Fiber Bragg gratings (FBGs) inscribed by UV light and different femtosecond laser techniques (phase mask, point-by-point, and plane-by-plane) were exposed-in several irradiation cycles-to accumulated high doses of gamma rays (up to 124 MGy) and neutron fluence (8.7 × 10/cm) in a research-grade nuclear reactor. The FBG peak wavelengths were measured continuously in order to monitor radiation-induced shifts.

View Article and Find Full Text PDF

With technological development, ionizing radiation has found applications in numerous occupations. However, the determination and quantification of the damage resulting from exposure to it remains rather unclear, along with the damage to particular organs. The aim of this systematic review was to investigate the relationship between low-dose ionizing radiation (LDIR) in exposed workers and possible functional changes and cancer development in the thyroid gland.

View Article and Find Full Text PDF

The melon fly, , poses a severe threat to the country's agricultural productivity, particularly in the cultivation of cucurbitaceous crops. This study was conducted to determine the ideal irradiation dose to be used to set up a Sterile Insect Technique (SIT)-based strategy to control outbreaks in Sri Lanka. A colony was established and maintained under standard laboratory conditions.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!