Introduction: Inhaled nitric oxide (INO) is a selective pulmonary vasodilator delivered from compressed gas cylinders filled to 2,200 psig (137.8 bar) with 800 ppm of NO in a balance of nitrogen. NO is currently FDA-approved for use in term or near-term infants with hypoxemia and signs of pulmonary hypertension in the absence of cardiac disease. INO has also been shown to improve oxygenation in adults with refractory hypoxemia. Current doctrine precludes the use of NO during military aeromedical transport owing to the requirement for large compressed gas cylinders. We performed a bench evaluation of 2 delivery systems that create NO from room air without the need for pressurized cylinders.

Materials And Methods: We evaluated 2 portable nitric oxide INO generation systems (LungFit PH, Beyond Air Inc, Garden City, NJ and a prototype NO generator, Odic Inc, Littleton, MA) at ground level, 8,000, and 14,000 feet (2,437 and 4,267 meter) simulated altitude in an altitude chamber. The output from each device was injected into the inspiratory limb of the ventilator circuit that was attached to a test lung. A 731 ventilator (Zoll Medical, Chelmsford, MA) and T1 (Hamilton Medical, Reno, NV) were used employing 24 combinations of ventilator settings each repeated in duplicate. An INOmax DS IR was used to measure delivered INO and NO2 via a sampling line attached in the ventilator circuit inspiratory limb. A fast response oxygen analyzer (O2CAP, Oxigraf Inc, Sunnyvale, CA) was used to measure inspired FiO2. Target INO concentration was 20 ppm.

Results: Across all ventilator settings, the LungFit device delivered INO was 19.8 ± 1.6 ppm, 16.1 ± 1.9 ppm, and 11.6 ± 1.7 ppm at ground level, 8,000 ft (2,437 meter), and 14,000 ft (4,267 meter), respectively. The Odic device delivered INO dose was 20.6 ± 1.4 ppm, 21.3 ± 5.5 ppm, and 20.4 ± 9.1 ppm at ground level, 8,000 ft (2,437 meter), and 14,000 ft (4,267 meter), respectively.

Conclusions: Both devices delivered a reliable INO dose at ground level. Altitude significantly affected INO delivery accuracy at 14,000 ft (4,267 meter) (P < 0.01) with both devices and at 8,000 ft (2,437 meter) (P < 0.01) with LungFit. Differences in INO dosage were not statistically significant with the Odic device at 8,000 ft (2,437 meter)(P > 0.05) although there were large variations with selected ventilator settings. With careful monitoring, devices creating INO from room air without cylinders could be used during aeromedical transport without the need for pressurized cylinders.

Download full-text PDF

Source
http://dx.doi.org/10.1093/milmed/usae195DOI Listing

Publication Analysis

Top Keywords

ground level
16
nitric oxide
12
delivered ino
12
14000 ft 4267 meter
12
ino
9
inhaled nitric
8
generation systems
8
oxide ino
8
compressed gas
8
gas cylinders
8

Similar Publications

The complexity of home-based rehabilitation technology implementation for post-stroke motor rehabilitation in the Netherlands.

BMC Health Serv Res

January 2025

Socio-Medical Sciences Department, Erasmus School of Health Policy & Management, Erasmus University Rotterdam, Rotterdam, The Netherlands.

Background: Rehabilitation technology is a growing field, but the sustainable implementation of these technologies, particularly in home settings, is lacking. The aim of this study was to explore the factors influencing the uptake of stroke rehabilitation technology among various stakeholders, including developers, healthcare professionals, individuals who had strokes, strategic experts, management and innovation staff, health insurers, and the National Health Care Institute.

Methods: In total, 22 semi-structured interviews were conducted with a purposive stakeholder sample.

View Article and Find Full Text PDF

Peroxidized lipids have been shown to reduce broiler performance whereupon it was theorized that dietary peroxide value (PV) plus anisidine value (AnV) may be predictive of broiler performance. In experiment (EXP) 1, 64 pens (8 broilers/pen) were fed diets containing 8 levels of peroxidized soybean oil (SO). Broilers were fed diets from 7 to 35 d of age with 8 replications per dietary treatment.

View Article and Find Full Text PDF

Granite sludge dust (GSD), a significant byproduct of granite processing globally, poses severe environmental and public health challenges, with India alone generating 200 million tons annually. The conventional use of GSD in soil stabilization and construction materials is limited to 20-30%, underscoring the urgent need for sustainable repurposing solutions within the circular economy catering to broader bulk utilization. Unlike traditional techniques, repurposing granite dust using microbially induced calcite precipitation (MICP) offers a sustainable low-impact and eco-friendly ground improvement solution.

View Article and Find Full Text PDF

Introduction: Virtual reality (VR) is a rapidly evolving technology that has been shown to improve pain severity in different disease states, including cancer. To date, VR pain studies have used off-the-shelf products for pain distraction. What are user preferences for VR content to mitigate cancer pain?.

View Article and Find Full Text PDF

Rationale And Objectives: Mixed ground-glass nodules (mGGNs) are highly malignant and common nonspecific lung imaging findings. This study aimed to explore whether combining quantitative and qualitative spectral dual-layer detector-based computed tomography (SDCT)-derived parameters with serological tumor abnormal proteins (TAPs) and thymidine kinase 1 (TK1) expression enhances invasive mGGN diagnostic efficacy and to develop a joint diagnostic model.

Materials And Methods: This prospective study included patients with mGGNs undergoing preoperative triple-phase contrast-enhanced SDCT with TAP and TK1 tests.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!