Aim: To invastigate intra- and interobserver reliability of interferometry, tear meniscus height (TMH) measurement and meibography (MBG) of an ocular surface analyzer, LacryDiag (Quantel Medical, France).

Methods: Five consecutive measurements and subsequent analysis of interferometry, TMH, and MBG were recorded by two examiners using the LacryDiag. To assess intra- and interobserver reliability, we used Cohen's kappa for categorical variables (interferometry), or intraclass correlation coefficient for continuous variables (TMH, MBG).

Results: Thirty eyes of 30 examinees were included. For both observers, there was excellent intraobserver reliability for MBG (0.955 and 0.970 for observer 1 and 2, respectively). Intraobserver reliability for observer 1 was substantial for interferometry (0.799), and excellent for TMH (0.863). Reliability for observer 2 was moderate for interferometry (0.535) and fair to good for TMH (0.431). Interobserver reliability was poor for interferometry (0.074) and fair to good for TMH (0.680) and MBG (0.414).

Conclusion: LacryDiag ocular surface analyzer in our study proves to be a reliable noninvasive tool for the evaluation of TMH and MBG. As for interferometry, poor interobserver reliability, fair to good intraobserver reliability for observer 1, and moderate for observer 2, leave room for improvement.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11286434PMC
http://dx.doi.org/10.18240/ijo.2024.08.05DOI Listing

Publication Analysis

Top Keywords

interobserver reliability
20
intra- interobserver
12
ocular surface
12
surface analyzer
12
intraobserver reliability
12
reliability observer
12
fair good
12
reliability
9
analyzer lacrydiag
8
tmh mbg
8

Similar Publications

Purpose: No studies have explored the reliability of the Rigo classification system using surface topography (ST), which would allow optimization without radiation exposure. This study aims to measure and compare the intra- and inter-observer reliability (Kappa values) and accuracy of the Rigo system between ST and X-ray for overall types and subtypes.

Methods: X-ray and ST images of 31 adolescent idiopathic scoliosis patients were selected.

View Article and Find Full Text PDF

Purpose: To propose a new sign of patellar maltracking in recurrent patellar dislocation (RPD) and compare the differences in lower limb rotational and bony structural abnormalities among the different signs.

Patients And Methods: A retrospective study included 279 patients (mean age: 22 years; female: 81%) who underwent primary surgery for RPD over the past 4 years was performed. The patients were grouped based on the characteristics of patellar tracking: low-, moderate- and high-grade J-sign.

View Article and Find Full Text PDF

Objectives: Multiple techniques have been proposed for functional fetal cardiology, including pulsed-wave (PW) and tissue Doppler imaging (TDI), Myocardial Performance Index (MPI), annular plane systolic excursion (TAPSE/MAPSE) and spatiotemporal image correlation (STIC). We aimed to compare these techniques' achievability and reproducibility to determine their clinical utility for each cardiac side.

Methods: Uncomplicated pregnancies from 22 to 39 weeks were recruited and images and volumes stored for offline analysis.

View Article and Find Full Text PDF

Node Reporting and Data System 1.0 (Node-RADS) for the Assessment of Oncological Patients' Lymph Nodes in Clinical Imaging.

J Clin Med

January 2025

Radiology, Multizonal Unit of Rovereto and Arco, APSS Provincia Autonoma Di Trento, 38123 Trento, Italy.

The assessment of lymph node (LN) involvement with clinical imaging is a key factor in cancer staging. Node Reporting and Data System 1.0 (Node-RADS) was introduced in 2021 as a new system specifically tailored for classifying and reporting LNs on computed tomography (CT) and magnetic resonance imaging scans.

View Article and Find Full Text PDF
Article Synopsis
  • Cutaneous T-cell lymphoma (CTCL) is a complex skin cancer that includes Mycosis fungoides and Sézary syndrome, making accurate diagnosis and severity assessment essential for effective treatment.
  • A study involving 16 dermatology residents showed significant variability in their evaluation of lesions using the mSWAT scoring system, particularly with tumors and lesions in erythrodermic patients, which were often misclassified.
  • The findings reveal the need for better training and standardized protocols in scoring to enhance reliability in assessing CTCL severity, similar to other assessment tools in dermatology.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!