Introduction: Percutaneous microwave ablation (MWA) is clinically accepted for the treatment of lung tumors and oligometastatic disease. Bronchoscopic MWA is under development and evaluation in the clinical setting. We previously reported on the development of a bronchoscopy-guided MWA system integrated with clinical virtual bronchoscopy and navigation and demonstrated the feasibility of transbronchial MWA, using a maximum power of 60 W at the catheter input. Here, we assessed the performance of bronchoscopy-guided MWA with an improved catheter (maximum power handling of up to 120 W) in normal porcine lung in vivo (as in the previous study).

Methods: A total of 8 bronchoscopy-guided MWA were performed ( = 2 pigs; 4 ablations per pig) with power levels of 90 W and 120 W applied for 5 and 10 min, respectively. Virtual bronchoscopy planning and navigation guided transbronchial or endobronchial positioning of the MWA applicator for ablation of lung parenchyma. Following completion of ablations and post-procedure CT imaging, the lungs were harvested and sectioned for gross and histopathologic ablation analysis.

Results: Bronchoscopy-guided MWA with applied energy levels of 90 W/5 min and 120 W/10 min yielded ablation zones with short-axis diameters in the range of 20-28 mm (56-116% increase) as compared to ∼13 mm from our previous study (60 W/10 min). Histology of higher-power and previous lower-power ablations was consistent, including a central necrotic zone, a thermal fixation zone with intact tissue architecture, and a hemorrhagic periphery. Catheter positioning and its confirmation via intra-procedural 3D imaging (e.g., cone-beam CT) proved to be critical for ablation consistency.

Conclusion: Bronchoscopy-guided MWA with an improved catheter designed for maximum power 120 W yields large ablations in normal porcine lung in vivo.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11324231PMC
http://dx.doi.org/10.1159/000539864DOI Listing

Publication Analysis

Top Keywords

bronchoscopy-guided mwa
20
porcine lung
12
maximum power
12
mwa
9
microwave ablation
8
virtual bronchoscopy
8
mwa improved
8
improved catheter
8
normal porcine
8
lung vivo
8

Similar Publications

Background: To treat multiple bilateral ground-glass opacities (GGOs), surgical treatments and electromagnetic navigation bronchoscopy (ENB)-guided ablation therapy are recommended therapeutic measures. However, the differences between bilateral and unilateral ablation, with or without surgery, remain unknown. This study aims to evaluate the differences in efficacy among various strategies.

View Article and Find Full Text PDF

Bronchoscopy-Guided High-Power Microwave Ablation in an in vivo Porcine Lung Model.

Biomed Hub

July 2024

Mike Wiegers Department of Electrical and Computer Engineering, Kansas State University, Manhattan, KS, USA.

Introduction: Percutaneous microwave ablation (MWA) is clinically accepted for the treatment of lung tumors and oligometastatic disease. Bronchoscopic MWA is under development and evaluation in the clinical setting. We previously reported on the development of a bronchoscopy-guided MWA system integrated with clinical virtual bronchoscopy and navigation and demonstrated the feasibility of transbronchial MWA, using a maximum power of 60 W at the catheter input.

View Article and Find Full Text PDF

Background: Patients with early-stage lung cancer are sometimes medically inoperable, and for patients with multiple primary lung cancers, surgical resection alone sometimes proves to be impractical. Local treatments like microwave ablation (MWA) are investigational alternatives for these patients. Most reported MWA procedures for lung cancers are performed percutaneously under CT guidance.

View Article and Find Full Text PDF

Here, we describe a novel method using microwave ablation (MWA) guided by electromagnetic navigation bronchoscopy (ENB) and video-assisted thoracoscopic surgery (VATS) for simultaneous treatment of multiple high-risk pulmonary nodules in a 47-year-old woman. After the ENB registration process, the operator delivered the locatable electromagnetic probe to the target in the right upper lobe following the navigational route. MWA was performed after an antenna was passed into the lesion through the working channel.

View Article and Find Full Text PDF

Background: Transbronchial lung biopsy is an important approach to diagnose peripheral lung cancer, but bronchoscopy based treatment options are limited and poorly studied. A flexible bronchoscopy-guided water-cooled microwave ablation (MWA) catheter was developed to evaluate the feasibility and safety both in and porcine models.

Methods: Using direct penetration of the catheter through the surface of porcine lung, ablations (n=9) were performed at 70, 80, 90 W for 10 minutes.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!