Purpose: Respiratory motion (RM) significantly impacts image quality in thoracoabdominal PET/CT imaging. This study introduces a unified data-driven respiratory motion correction (uRMC) method, utilizing deep learning neural networks, to solve all the major issues caused by RM, i.e., PET resolution loss, attenuation correction artifacts, and PET-CT misalignment.

Methods: In a retrospective study, 737 patients underwent [F]FDG PET/CT scans using the uMI Panorama PET/CT scanner. Ninety-nine patients, who also had respiration monitoring device (VSM), formed the validation set. The remaining data of the 638 patients were used to train neural networks used in the uRMC. The uRMC primarily consists of three key components: (1) data-driven respiratory signal extraction, (2) attenuation map generation, and (3) PET-CT alignment. SUV metrics were calculated within 906 lesions for three approaches, i.e., data-driven uRMC (proposed), VSM-based uRMC, and OSEM without motion correction (NMC). RM magnitude of major organs were estimated.

Results: uRMC enhanced diagnostic capabilities by revealing previously undetected lesions, sharpening lesion contours, increasing SUV values, and improving PET-CT alignment. Compared to NMC, uRMC showed increases of 10% and 17% in SUV and SUV across 906 lesions. Sub-group analysis showed significant SUV increases in small and medium-sized lesions with uRMC. Minor differences were found between VSM-based and data-driven uRMC methods, with the SUV was found statistically marginal significant or insignificant between the two methods. The study observed varied motion amplitudes in major organs, typically ranging from 10 to 20 mm.

Conclusion: A data-driven solution for respiratory motion in PET/CT has been developed, validated and evaluated. To the best of our knowledge, this is the first unified solution that compensates for the motion blur within PET, the attenuation mismatch artifacts caused by PET-CT misalignment, and the misalignment between PET and CT.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11599311PMC
http://dx.doi.org/10.1007/s00259-024-06872-xDOI Listing

Publication Analysis

Top Keywords

respiratory motion
16
urmc
9
motion
8
resolution loss
8
loss attenuation
8
attenuation correction
8
correction artifacts
8
artifacts pet-ct
8
pet-ct misalignment
8
data-driven respiratory
8

Similar Publications

This study was conducted to evaluate the use of 4-dimensional (4D) maximum intensity projection (4D-MIP) to compensate for the disadvantages of average intensity projection (AIP), which is used to determine the internal target volume (ITV) in lung tumors. A respiratory motion phantom with a simulated tumor was imaged using 4D computed tomography (4D-CT). AIP and 4D-MIP were generated based on 10 phases of 4D-CT, followed by contouring of the ITV and ITV; these were compared with the ITV contoured in 10 phases of 4D-CT (ITV).

View Article and Find Full Text PDF

Continuous respiration monitoring is an important tool in assessing the patient's health and diagnosing pulmonary, cardiovascular, and sleep-related breathing disorders. Various techniques and devices, both contact and contactless, can be used to monitor respiration. Each of these techniques can provide different types of information with varying accuracy.

View Article and Find Full Text PDF

Multispectral transmission imaging has emerged as a promising technique for imaging breast tissue with high resolution. However, the method encounters challenges such as low grayscale, noisy transmission images with weak signals, primarily due to the strong absorption and scattering of light in breast tissue. A common approach to improve the signal-to-noise ratio (SNR) and overall image quality is frame accumulation.

View Article and Find Full Text PDF

Whole-body PET imaging is often hindered by respiratory motion during acquisition, causing significant degradation in the quality of reconstructed activity images. An additional challenge in PET/CT imaging arises from the respiratory phase mismatch between CT-based attenuation correction and PET acquisition, leading to attenuation artifacts. To address these issues, we propose two new, purely data-driven methods for the joint estimation of activity, attenuation, and motion in respiratory self-gated TOF PET.

View Article and Find Full Text PDF

Posterior sternoclavicular joint (SCJ) dislocation is a rare but potentially life-threatening injury due to its proximity to critical mediastinal structures. Early diagnosis and prompt management are essential to prevent severe complications such as vascular or respiratory compromise. We report a case of a 23-year-old male who presented to our emergency department five days after a high-energy motor vehicle accident with isolated, closed posterior dislocation of the SCJ.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!