Introduction: Assessment of left ventricular ejection fraction (LVEF) is one of the primary objectives of echocardiography. The gold standard assessment technique in emergency medicine is eyeballing. A new tool is now available on pocket ultrasound devices (PUD): automatic LVEF. The primary aim of this study was to evaluate the concordance between LVEF values estimated by automatic LVEF with PUD and by cardiac magnetic resonance imaging (MRI).

Materials: This was a prospective, monocentric, and observational study. All adult patients with an indication for cardiac MRI underwent a point-of-care ultrasound. Blinded to the MRI results, the emergency physician assessed LVEF using the automatic PUD tool and by visual evaluation.

Results: Sixty patients were included and analyzed. Visual estimation of LVEF was feasible for all patients and automatic evaluation for 52 (87%) patients. Lin's concordance correlation coefficient between automatic ejection fraction with PUD and by cardiac MRI was 0.23 (95% CI, 0.03-0.40).

Conclusion: Concordance between LVEF estimated by the automatic ejection fraction with PUD and LVEF estimated by MRI was non-existent.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11318925PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0308580PLOS

Publication Analysis

Top Keywords

ejection fraction
16
left ventricular
8
ventricular ejection
8
tool pocket
8
pocket ultrasound
8
cardiac magnetic
8
magnetic resonance
8
resonance imaging
8
lvef
8
lvef primary
8

Similar Publications

Heart remodelling affects ECG in rat DOCA/salt model.

Physiol Res

December 2024

Department of Physiology, Faculty of Medicine, Masaryk University, Brno, Czech Republic.

Myocardial remodelling involves structural and functional changes in the heart, potentially leading to heart failure. The deoxycorticosterone acetate (DOCA)/salt model is a widely used experimental approach to study hypertension-induced cardiac remodelling. It allows to investigate the mechanisms underlying myocardial fibrosis and hypertrophy, which are key contributors to impaired cardiac function.

View Article and Find Full Text PDF

Multisociety endorsement of the 2024 European guideline recommendations on coronary revascularization.

J Thorac Cardiovasc Surg

January 2025

Coronary Center, Department of Thoracic and Cardiovascular Surgery, Miller Family Heart, Vascular, & Thoracic Institute, Cleveland Clinic, Cleveland, Ohio. Electronic address:

View Article and Find Full Text PDF

Unmasking High-Output Heart Failure: Beyond Conventional Paradigms.

Cardiol Rev

January 2025

From the Department of Internal Medicine, Division of Cardiology, Wayne State University, Detroit, MI.

Heart failure (HF) poses a significant medical challenge, affecting millions of adults in the United States. High-output heart failure (HOHF) is a distinct subtype characterized by elevated cardiac output exceeding 8 L/min or a cardiac index >4 L/min/m². Patients with HOHF often present similarly to those with heart failure with reduced ejection fraction and heart failure with preserved ejection fraction.

View Article and Find Full Text PDF

Evaluating cardiac performance in beagle dogs: Transesophageal echocardiography and myocardial work assessment.

Heliyon

January 2025

Department of Ultrasound Medicine, Nanjing Drum Tower Hospital, Affiliated Hospital of Medical School, Nanjing University, Nanjing, 210008, People's Republic of China.

Objectives: This study aimed to establish standard transesophageal echocardiographic (TEE) measurements of left ventricular (LV) morphology, function, and myocardial work parameters in healthy Beagle dogs using pressure-strain loops (PSL). Additionally, it sought to standardize optimal TEE imaging techniques and explore the potiential application of myocardial work analyis in veterinary medicine.

Methods: Thirty-seven healthy male Beagle dogs were anesthetized, intubated, and mechanically ventilated for TEE examinations.

View Article and Find Full Text PDF

Cardiovascular and respiratory alterations during anesthesia are of major concern in canines. Thus, it is essential to understand the potential depressant effects of anesthetic drugs on cardio-vascular system; so that, anesthetic procedures are conducted in the best possible way. The objective of the study was to assess and compare the echocardiographic indices during dex-medetomidine and midazolam anesthesia in dogs undergoing elective ovariohysterectomy.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!