Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Outcomes in patients with relatively high His-bundle (HB) capture thresholds at implantation are unknown. This study aimed to compare changes in the HB capture threshold and prognosis between patients with a relatively high threshold and those with a low threshold.
Methods And Results: Forty-nine patients who underwent permanent HB pacing (HBP) were divided into two groups: low (<1.25 V at 1.0 ms; n=35) and high (1.25-2.49 V; n=14) baseline HB capture threshold groups. The HB capture threshold was evaluated at implantation, and after 1 week, 1, 3, and 6 months, and every 6 months thereafter. HB capture threshold rise was defined as threshold rise ≥1.0 V at 1.0 ms compared with implantation measures. We compared outcomes between the groups. During a mean follow-up period of 34.6 months, the high-threshold group showed a trend toward a higher incidence of HB capture threshold of ≥2.5 V (50% vs. 14%; P=0.023), HBP abandonment (29% vs. 8.6%; P=0.091), lead revision (21% vs. 2.9%; P=0.065), and clinical events (all-cause death, heart failure hospitalization, and new-onset or progression of atrial fibrillation; 50% vs. 23%; P=0.089) than the low-threshold group. A baseline HB capture threshold of ≥1.25V was an independent predictor of clinical events.
Conclusions: A relatively high HB capture threshold is associated with increased risk of HBP abandonment, lead revision, and poor clinical outcomes.