Background: The PIONEER III trial demonstrated noninferiority of 12-month target lesion failure (TLF) with the Supreme DES (Sinomed), a thin-strut cobalt-chromium, biodegradable polymer, sirolimus-eluting stent, compared with a durable polymer, everolimus-eluting (XIENCE/PROMUS) stent (DP-EES). The relative safety and effectiveness of the Supreme DES in patients with acute coronary syndromes (ACS) and those with chronic coronary syndromes (CCS) is not known.

Methods: PIONEER III was a prospective, multicenter, international, 2:1 randomized trial stratified by clinical presentation. The primary end point was TLF at 12 months (a composite of cardiac death, target vessel myocardial infarction, or ischemia-driven target lesion revascularization).

Results: A total of 1628 patients were enrolled, including 41% of patients with ACS (unstable angina and non-ST-elevation myocardial infarction) randomized to Supreme DES (n = 441) versus DP-EES (n = 232) and 59% of patients with CCS randomized to Supreme DES (n = 645) versus DP-EES (n = 310). Patients with ACS were younger, fewer presented with less diabetes, hypertension, and previous revascularization, but more were current smokers. The primary end point of TLF (6.4% vs 4.4%; = .1), major adverse cardiac events (8.5% vs 6.5%; = .16), and stent thrombosis (0.4% vs 0.9%; = .25) at 12 months were similar in the ACS and CCS groups. There was no difference in TLF at 12 months between Supreme DES and DP-EES among patients with ACS (6.6% vs 6.0%; = .89) and those with CCS (4.5% vs 4.3%; = .83); interaction = .51 for TLF by clinical presentation.

Conclusions: Compared with the DP-EES, the Supreme DES seemed safe and effective with a similar TLF at 12 months in both patients with ACS and those with CCS.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11307614PMC
http://dx.doi.org/10.1016/j.jscai.2023.100629DOI Listing

Publication Analysis

Top Keywords

supreme des
24
patients acs
16
coronary syndromes
12
pioneer iii
12
tlf months
12
biodegradable polymer
8
chronic coronary
8
target lesion
8
primary point
8
point tlf
8

Similar Publications

Background: The PIONEER III trial demonstrated noninferiority of 12-month target lesion failure (TLF) with the Supreme DES (Sinomed), a thin-strut cobalt-chromium, biodegradable polymer, sirolimus-eluting stent, compared with a durable polymer, everolimus-eluting (XIENCE/PROMUS) stent (DP-EES). The relative safety and effectiveness of the Supreme DES in patients with acute coronary syndromes (ACS) and those with chronic coronary syndromes (CCS) is not known.

Methods: PIONEER III was a prospective, multicenter, international, 2:1 randomized trial stratified by clinical presentation.

View Article and Find Full Text PDF

Significant climate finance gaps exist for small islands in transitioning to net zero, as climate commitments far outweigh the government budget and international financing. To create alignment between resource allocation and climate commitments, a roadmap for strategic and cost-effective decarbonization is of supreme importance. This paper presents a geographic carbon accounting model which incorporates emissions from electricity, transportation, food systems, and human respiration, whilst accounting for the carbon uptake by the terrestrial biosphere in view of identifying high-intensity aggregated emissions estimated in the range of 200-215 ktCO in the coastal and inland urban regions of the remote island of Mauritius.

View Article and Find Full Text PDF

Background: The PIONEER III trial showed the 12-month safety and efficacy of the Supreme drug-eluting stent (DES) vs the durable polymer everolimus-eluting stent. We sought to assess whether the characteristics and clinical outcomes of the Supreme DES in PIONEER III were consistent among patients by enrollment location.

Methods: This subgroup analysis of the PIONEER III trial compared the characteristics and outcomes of patients recruited from North America, Europe, and Japan and the relative differences in patient outcomes according to the site recruitment volume.

View Article and Find Full Text PDF

Background: Patients with diabetes mellitus (DM) have worse outcomes following percutaneous coronary intervention than nondiabetic patients. The novel Supreme DES is a biodegradable polymer sirolimus-eluting stent designed to synchronize early drug delivery, limiting the potential for long-term inflammatory response. The purpose of this study was to evaluate the safety and efficacy of the Supreme DES in patients with DM.

View Article and Find Full Text PDF

Background: The Supreme healing-targeted drug-eluting stent (DES) is designed to promote endothelial healing to reduce stent-related adverse events. This may be particularly relevant among complex lesions that have a higher rate of adverse events. We sought to compare 1-year outcomes of percutaneous coronary intervention in complex lesions between the Supreme DES and contemporary durable-polymer, everolimus-eluting stents (DP-EES).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!