Introduction: Laboratories use their performance in external quality assurance (EQA) to establish quality planning strategies and to assess whether testing processes require improvement.
Methods: The EQA performance of the hematology and coagulation test parameters on the Royal College of Pathologists of Australasia EQA program was evaluated over a 4-year cycle at an academic hospital in Johannesburg, South Africa. The test performance was determined from analytical quality specification (APS) and/or z-scores. Bias and imprecision were used to calculate sigma (σ) metric scores. Specifications from European Federation of Laboratory Medicine and/or biological variation were applied.
Results: The laboratory achieved a mean testing score of 98.7 ± 4.0%. There were 103 (10.7%) unacceptable results. On investigation, root causes included: presurvey issues (83%), transcription errors (9%), random errors (6%), and test performance errors (3%). All test parameters evaluated achieved an acceptable median APS during the study period. The mean z-scores, however, were >2 and unacceptable for mean cell hemoglobin concentration and hematocrit. On investigation, this was attributed to significant delay in transport and storage of full blood count samples. White cell count and d-dimer achieved a σ ≥ 6.
Conclusion: EQA participation assisted the laboratory in maintaining a quality system. Close monitoring is necessary for international laboratories to avoid sample delays that can affect result quality.