Objectives: 7α-Hydroxy-4-cholesten-3-one (C4) is the common intermediary of both primary bile acids. C4 is recommended by the British Society of Gastroenterology for the investigation of bile acid diarrhoea (BAD) in patients with chronic diarrhoea. This project aimed to develop and validate an assay to quantitate C4 in serum and assess the stability of C4 in unseparated blood.
Methods: Accuracy was underpinned by calibrating to quantitative nuclear magnetic resonance analysis. C4 was analysed in a 96-well plate format with a deuterated C4 internal standard and liquid-liquid extraction. Validation followed the 2018 Food and Drug Administration guidelines. To assess C4 stability, healthy volunteers (n=12) donated 8 fasted samples each. Samples were incubated at 20 °C for up to 72 h and retrieved, centrifuged, aliquoted and frozen for storage at different time points prior to C4 analysis.
Results: The C4 method demonstrated excellent analytical performance and passed all validation criteria. The method was found to be accurate, precise, free from matrix effects and interference. After 72 h of delayed sample separation, C4 concentration gradually declined by up to 14 % from baseline. However, the change was not significant for up to 12 h.
Conclusions: We present a robust method of analysing serum C4, offering a convenient alternative to SeHCAT for BAD investigation. C4 was found to decline in unseparated blood over time; however, after 12 h the mean change was <5 % from baseline. Our results suggest C4 is suitable for collection from both primary and secondary care prior to gastroenterology referral.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1515/cclm-2024-0275 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!