Purpose: To ensure comparable PET/CT image quality between or within centres, clinical inter-system performance comparisons following European Association of Nuclear Medicine Research Ltd. (EARL) guidelines is required. In this work the performance of the long axial field-of-view Biograph Vision Quadra is compared to its predecessor, the short axial field-of-view Biograph Vision.

Procedures: To this aim, patients with suspected tumour lesions received a single weight-based (3 MBq/kg) 2-deoxy-2-[F]fluoro-D-glucose injection and underwent routine clinical ( 15 min) scans on the Vision and 3-min scans on the Quadra in listmode in balanced order. Image quality (IQ), image noise (IN), and tumour demarcation (TD) were assessed visually by four nuclear medicine physicians using a 5-point Likert scale and semiquantitative analysis was performed using standardised uptake values (SUVs). Inter-reader agreement was tested using Wilcoxon's signed rank test and the SUVs were statistically compared using a paired t-test.

Results: Twenty patients (mean age, 60 years ± 8.8 [standard deviation], 16 male) were enrolled. Inter-reader agreement ranged from good to very good for IQ and IN (0.62 ≤ W ≤ 0.81), and fair for TD (0.29 ≤ W ≤ 0.39). Furthermore, a significant difference was found for TD (p = 0.015) between the systems, showing improved TD for the Quadra.

Conclusion: This study demonstrates that the Quadra can be used in routine clinical practice with multiple PET/CT systems or in multicentre studies. This system provides comparable diagnostic image quality and semiquantitative accuracy, improved TD, and has the advantage of shorter scan durations.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11436434PMC
http://dx.doi.org/10.1007/s11307-024-01939-5DOI Listing

Publication Analysis

Top Keywords

axial field-of-view
12
image quality
12
short axial
8
nuclear medicine
8
field-of-view biograph
8
routine clinical
8
inter-reader agreement
8
clinical
4
clinical performance
4
performance comparison
4

Similar Publications

Objective: Timing calibration is essential for positron emission tomography (PET) system as it enhances timing resolution to improve image quality. Traditionally, positron sources are employed for timing calibration. However, the photons emitted by these sources travel in opposite directions, necessitating that positrons annihilate at multiple locations to collect coincidence data across a greater number of lines of response (LORs).

View Article and Find Full Text PDF

High-sensitivity total-body PET enables faster scans, lower doses, and dynamic multiorgan imaging. However, the higher system cost of a scanner with a long axial field of view (AFOV) hinders its wider application. This paper investigates the impact on the lesion quantification and detectability of cost-effective total-body PET sparse designs.

View Article and Find Full Text PDF

Introduction: Mu-opioid receptors (MORs) are G-coupled protein receptors with a high affinity for both endogenous and exogenous opioids. MORs are widely expressed in the central nervous system (CNS), peripheral organs, and the immune system. They mediate pain and reward and have been implicated in the pathophysiology of opioid, cocaine, and other substance use disorders.

View Article and Find Full Text PDF

: Paediatric PET/CT imaging is crucial in oncology but poses significant radiation risks due to children's higher radiosensitivity and longer post-exposure life expectancy. This study aims to minimize radiation exposure by generating synthetic CT (sCT) images from emission PET data, eliminating the need for attenuation correction (AC) CT scans in paediatric patients. : We utilized a cohort of 128 paediatric patients, resulting in 195 paired PET and CT images.

View Article and Find Full Text PDF

Waveguide evanescent field fluorescence microscopy (WEFF) is an evanescent-based microscopy that utilizes a confined thin film of light, around 100 nm, to image the plasma membrane of cells attached to a waveguide. Low photobleaching and low background besides its high axial resolution allows time-lapse imaging to investigate changes in cell morphology in the presence or absence of chemical agents. Both large field of view (FOV) and uniform illumination are very important while imaging cell-substrate contacts with an evanescent field.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!