A Faraday-effect radial interferometer-polarimeter is designed for the National Spherical Torus Experiment-Upgrade (NSTX-U) to measure multiscale magnetic and density fluctuations critical to understanding fusion plasma confinement and stability, including those originating from magnetohydrodynamic instabilities, energetic particle-driven modes, and turbulence. The diagnostic will utilize the three-wave technique with 5 MHz bandwidth to simultaneously measure line-integrated magnetic and density fluctuations up to the ion-cyclotron frequency. Probe beams will be launched radially from the low-field side at the NSTX-U midplane, where the measured Faraday fluctuations mainly correspond to radial magnetic fluctuations that directly link to magnetic transport. A correlation technique will be employed to reduce the measurement noise to below 0.01° enabling detection of small amplitude fluctuations. Two toroidally displaced chords with 7° separation will be installed to measure toroidal mode numbers up to n = 25 for mode identification. Solid-state microwave sources operating at 321 μm (935 GHz) will be used to minimize the impact of the Cotton-Mouton effect.

Download full-text PDF

Source
http://dx.doi.org/10.1063/5.0217322DOI Listing

Publication Analysis

Top Keywords

magnetic density
12
radial interferometer-polarimeter
8
national spherical
8
spherical torus
8
torus experiment-upgrade
8
experiment-upgrade nstx-u
8
density fluctuations
8
magnetic
5
fluctuations
5
will
5