AI Article Synopsis

  • The study investigates the impact of different types of total knee arthroplasty (TKA)—cruciate-retaining (CR) and posterior-stabilized (PS)—on biomechanical characteristics in cadaver knees under varying weight-bearing conditions.
  • The results indicate that CR implants experience greater posterior femur translation at higher flexion angles compared to PS implants, while PS implants generate a higher patellofemoral contact force.
  • At specific flexion angles, CR implants show significantly more tibiofemoral contact force than PS implants, suggesting different mechanical behaviors between the two designs under load.

Article Abstract

Background: Increased load bearing across the patellofemoral and tibiofemoral articulations has been associated with total knee arthroplasty (TKA) complications. Therefore, the purpose of this study was to quantify the biomechanical characteristics of the patellofemoral and tibiofemoral joints and simulate varying weight-bearing demands after posterior cruciate ligament-retaining (CR) and posterior-stabilized (PS) TKAs.

Methods: Eight fresh-frozen cadaveric knees (average age, 68.4 years; range, 40-86 years) were tested using a custom knee system with muscle-loading capabilities. The TKA knees were tested with a CR and then a PS TKA implant and were loaded at 6 different flexion angles from 15° to 90° with progressively increasing loads. The independent variables were the implant types (CR and PS TKA), progressively increased loading, and knee flexion angle (KFA). The dependent variables were the patellofemoral and tibiofemoral kinematics and contact characteristics.

Results: The results showed that at higher KFAs, the position of the femur translated significantly more posterior in CR implants than in PS implants (36.6 ± 5.2 mm and 32.5 ± 5.7 mm, respectively). The patellofemoral contact force and contact area were significantly greater in PS than in CR implants at higher KFAs and loads (102.4 ± 12.5 N and 88.1 ± 10.9 N, respectively). Lastly, the tibiofemoral contact force was significantly greater in the CR than the PS implant at flexion angles of 45°, 60°, 75°, and 90° KFA, the average at these flexion angles for all loads tested being 246.1 ± 42.1 N and 192.8 ± 54.8 N for CR and PS implants, respectively.

Conclusions: In this biomechanical study, CR TKAs showed less patellofemoral contact force, but more tibiofemoral contact force than PS TKAs. For higher loads across the joint and at increased flexion angles, there was significantly more posterior femur translation in the CR design with a preserved posterior cruciate ligament and therefore significantly less patellofemoral contact area and force than in the PS design. The different effects of loading on implants are an important consideration for physicians as patients with higher load demands should consider the significantly greater patellofemoral contact force and area of the PS over the CR design.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11262949PMC
http://dx.doi.org/10.4055/cios23356DOI Listing

Publication Analysis

Top Keywords

contact force
20
flexion angles
16
patellofemoral contact
16
patellofemoral tibiofemoral
12
total knee
8
knee arthroplasty
8
biomechanical study
8
posterior cruciate
8
contact
8
higher kfas
8

Similar Publications

Knitted fabrics are metamaterials with remarkable mechanical properties, such as extreme deformability and multiple history-dependent rest shapes. This Letter shows that those properties may stem from a continuous set of metastable states for a fabric free of external forces. This is evidenced through experiments, numerical simulations, and analytical developments.

View Article and Find Full Text PDF

This study focuses on the fabrication and characterisation of single-walled carbon nanotube (SWCNT) buckypapers and polyethersulfone (PES) flat-sheet membranes using Cyrene, aiming toevaluate its efficacy as a green solvent for these applications. Pristine SWCNTs were dispersed inCyrene without surfactants and compared to N-Methyl-2-pyrrolidone (NMP) dispersions. Buckypapers were fabricated from these dispersions and characterised using Scanning ElectronMicroscopy (SEM), Atomic Force Microscopy (AFM), and infrared spectroscopy.

View Article and Find Full Text PDF

Background: For patients with osteoporosis and rotator cuff tears, there is still no consensus on current treatment methods. The material, structure, and number of anchors have important effects on the repair outcome.

Purpose: To investigate the use of chitosan quaternary ammonium salt-coated nickel-titanium memory alloy (NTMA) anchors to treat rotator cuff injury in shoulders with osteoporosis in a rabbit osteoporosis model.

View Article and Find Full Text PDF

Construction of Loop Polyzwitterion Brushes on PET Sheets via the Chain-End Closure Strategy To Improve Antifouling and Hemocompatible Properties.

Langmuir

January 2025

Jiangsu Collaborative Innovation Center of Biomedical Functional Materials, Jiangsu Key Laboratory of Bio-functional Materials, Department of Materials Science and Engineering, School of Chemistry and Materials Science, Nanjing Normal University, Nanjing 210023, P. R. China.

Steric stabilization and lubrication give loop polymer brushes enhanced antifouling properties. In the study, linear zwitterionic poly(NMASMCMS) brushes were first constructed on a poly(ethylene terephthalate) (PET) surface through surface-initiated reversible addition-fragmentation chain-transfer (SI-RAFT) polymerization. The tethered linear brushes on sheets were then thiolated with ethanolamine, followed by oxidation to form loop brushes.

View Article and Find Full Text PDF

In order to reduce turnout rail wear, the paper establishes a coupled dynamics model and a turnout rail wear model that consider the true profile of the turnout rail, the vehicle's continuous traction force while passing, and the operational resistance. Comparative analysis of various models for predicting turnout rail wear indicates that the wear energy model is better suited for this purpose. The ideal profile update step for the turnout rail is 0.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!