Objective: As the prevalence of neuroendovascular interventions increases, it is critical to mitigate unnecessary radiation for patients, providers, and health care staff. Our group previously demonstrated reduced radiation dose and exposure during diagnostic angiography by reducing the default pulse and frame rates. We applied the same technique for basic neuroendovascular interventions.
Methods: We performed a retrospective review of prospectively acquired data after implementing a quality improvement protocol in which pulse rate and frame rate were reduced from 15 p/s to 7.5 p/s and 7.5 f/s to 4.0 f/s respectively. We studied consecutive, unilateral middle meningeal artery embolizations treated with particles. Total radiation dose, radiation per angiographic run, total radiation exposure, and exposure per run were calculated. Multivariable log-linear regression was performed to account for patient body mass index (BMI), number of angiographic runs, and number of vessels catheterized.
Results: A total of 20 consecutive, unilateral middle meningeal artery embolizations were retrospectively analyzed. The radiation reduction protocol was associated with a 39.2% decrease in the total radiation dose and a 37.1% decrease in radiation dose per run. The protocol was associated with a 41.6% decrease in the total radiation exposure and a 39.5% decrease in exposure per run.
Conclusions: Radiation reduction protocols can be readily applied to neuroendovascular interventions without increasing overall fluoroscopy time and reduce radiation dose and exposure by 39.2% and 41.6% respectively. We strongly encourage all interventionalists to be cognizant of pulse rate and frame rate when performing routine interventions.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11449538 | PMC |
http://dx.doi.org/10.7461/jcen.2024.E2024.02.003 | DOI Listing |
Sci Rep
December 2024
Department of Cardiology, West China Hospital of Sichuan University, 37 Guoxue Alley, Wuhou District, Chengdu, 610041, Sichuan, China.
Intracardiac echocardiography (ICE) has been used to guide radio-frequency catheter ablation (RFCA) for better catheter navigation and less radiation exposure in treating atrial fibrillation (AF). This retrospective cohort study enrolled 227 AF patients undergoing ICE- or traditional fluoroscopy (TF)-guided RFCA for AF in a tertiary hospital. ICE was used more often in patients with atrial tachycardia [odds ratio (OR) 3.
View Article and Find Full Text PDFPituitary
December 2024
Department of Endocrinology and Nutrition, Hospital Universitario de Puerta de Hierro Majadahonda, Madrid, Spain.
Purpose: Studies focused on the effects of sellar and/or perisellar (S/PS) meningiomas on pituitary function are scarce. The primary objective of the present study was to determinate the effects that S/PS meningiomas and their treatments have on pituitary function. Also, we described the clinical characteristics and therapeutic outcomes of the cohort of adult Spanish patients.
View Article and Find Full Text PDFSci Rep
December 2024
Princess Margaret Cancer Centre, University Health Network, Toronto, Canada.
Despite decades of improvements in cytotoxic therapy, the current standard of care for locally advanced pancreatic cancer (LAPC) provides, on average, only a few months of survival benefit. Stereotactic Body Radiation Therapy (SBRT), a technique that accurately delivers high doses of radiation to tumors in fewer fractions, has emerged as a promising therapy to improve local control of LAPC; however, its effects on the tumor microenvironment and hypoxia remain poorly understood. To explore how SBRT affects pancreatic tumors, we combined an orthotopic mouse model of pancreatic cancer with an intravital microscopy platform to visualize changes to the in vivo tumor microenvironment in real-time.
View Article and Find Full Text PDFBiomed Chromatogr
January 2025
State Key Laboratory of Radiation Medicine and Protection, School of Radiation Medicine and Protection, Medical College of Soochow University, Collaborative Innovation Center of Radiological Medicine of Jiangsu Higher Education Institutions, Soochow University, Suzhou, China.
An animal model of radiation-induced lung injury (RILI) was established using female rats given sublethal whole-thorax X-ray irradiation (15 Gy) at a dose rate of 2.7 Gy/min. The rats were studied for up to day 45 and compared with sham-irradiated controls.
View Article and Find Full Text PDFJ Environ Radioact
December 2024
Radiation Epidemiology Branch, Division of Cancer Epidemiology and Genetics, National Cancer Institute, DHHS, NIH, Bethesda, MD, 20892-9778, USA; Faculty of Health and Life Sciences, Oxford Brookes University, Headington Campus, Oxford, OX3 0BP, UK. Electronic address:
We investigate methods that improve the estimation of indoor gamma ray dose rates at locations where measurements had not been made. These new predictions use a greater range of modelling techniques and larger variety of explanatory variables than our previous examinations of this subject. Specifically, we now employ three types of machine learning models in addition to the geostatistical, nearest neighbour and other earlier models.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!