Background And Aims: Transcatheter aortic valve replacement (TAVR) determined a paradigm shift in the treatment of patients with severe symptomatic aortic stenosis but data on very long-term durability are lacking. We sought to evaluate the clinical and hemodynamic outcomes of the CoreValve porcine pericardial self-expanding bioprosthesis at 12-year follow-up.

Methods: 882 inoperable or high-risk patients were treated with the CoreValve bioprosthesis in 8 Italian high-volume centers between 2007 and 2011. The endpoints were 12-year all-cause and cardiovascular mortality, and Cumulative Incidence Functions (CIFs) for severe Structural Valve Deterioration (SVD), Bioprosthetic Valve Dysfunction (BVD), Bioprosthetic Valve Failure (BVF), and severe Hemodynamic Valve Deterioration (HVD). VARC-3 definitions were applied.

Results: Baseline characteristics included a mean age of 83 ± 6 years, and NYHA class III or IV in 76.3 % of patients. The actuarial risk of death at 12 years after TAVR was 95.5 % (CI 93.5 %- 97.1 %). The actual risk of cardiovascular death, weighted against the risk of non-cardiac death at 12 years was 23.9 % (21.0 %-26.8 %). The 12-year actual risk of BVD was 7.0 % (5.3 %-8.9 %), of SVD was 3.6 % (2.5 %-5.2 %), of BVF was 3.12 % (2.02 %-4.57 %), and of severe HVD was 1.7 % (0.9 %-2.9 %). Mean transaortic gradient significantly decreased after the procedure (52 ± 15 mmHg vs 9 ± 5 mmHg, p < 0.001), and remained stable up to 12 years (12 ± 4 mmHg, P = 0.08 vs. discharge).

Conclusions: The first-generation CoreValve bioprosthesis showed reassuring clinical and hemodynamic performance at 12-year follow-up.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.carrev.2024.07.013DOI Listing

Publication Analysis

Top Keywords

transcatheter aortic
8
aortic valve
8
valve replacement
8
self-expanding bioprosthesis
8
valve deterioration
8
bioprosthetic valve
8
death 12 years
8
actual risk
8
valve
6
replacement corevalve
4

Similar Publications

Background: Evidence regarding the incidence of prosthesis-patient mismatch (PPM) and long-term mortality after transcatheter aortic valve replacement (TAVR) in patients with bicuspid aortic valve stenosis (AS) is scarce.

Objectives: This study sought to assess the incidence and prognostic impact of PPM after TAVR for bicuspid AS compared with that for tricuspid AS.

Methods: In total, 7,393 patients who underwent TAVR were prospectively enrolled in the OCEAN-TAVI (Optimized Catheter Valvular Intervention Transcatheter Aortic Valve Implantation) registry, an ongoing Japanese, multicenter registry.

View Article and Find Full Text PDF

Severe mitral regurgitation (MR) following surgical repair of the mitral valve poses a significant clinical challenge. Patients who have undergone surgery are typically at high risk for a second operation. This report details the case of a 54-year-old male who underwent aortic valve replacement and mitral valve repair using a 34-ring, 14 years prior.

View Article and Find Full Text PDF

According to current guidelines, patients with heart valve disease should be followed by Heart Valve Clinics (HVCs). Regular quality analysis is a major prerequisite of an HVC's program, but few data have been reported so far. We retrospectively collected patients with isolated, native aortic valve stenosis who had been visited in our HVC at least once between 2021 and 2024.

View Article and Find Full Text PDF

Patients with end-stage kidney disease (ESRD) represent a high-risk population in terms of both development of and death by cardiovascular diseases. Outcome data of ESRD patients with severe aortic valve stenosis (AS) treated by transcatheter aortic valve implantation (AVI) are scarce. We aim to compare the outcome of ESRD patients undergoing transfemoral (TF) or transapical (TA) AVI.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!