As technology in mechanical circulatory support continues to advance, so does the utilization of durable left ventricular assist devices (LVADs). We present the first reported case of HeartMate 3 (HM3) LVAD bend relief disconnection causing laceration of the outflow graft. Our patient is a 34 year old male with end-stage heart failure status post-HM3 implantation with recent debridement of a driveline infection who presented with weakness and sanguineous drainage 5 days after an episode of heavy lifting associated with a palpable substernal pop. During the initial evaluation, he experienced sudden massive decompression of a left chest hematoma through a thoracotomy incision with exsanguinating hemorrhage and an abrupt drop in LVAD flow. An emergent anterolateral thoracotomy was performed where it was discovered that the bend relief was disconnected and had lacerated the outflow graft. The graft was repaired and the ring was removed using a diamond blade rotary saw. He was discharged home and made a full recovery. Mechanical device malfunction has been documented in several prior versions of durable LVADS. Although there have been reported cases in the Heartmate 2, this is the first reported case of a HM3 bend relief disconnection and resultant laceration of the outflow graft.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/MAT.0000000000002287 | DOI Listing |
J Clin Med
December 2024
Department of Cardiothoracic and Vascular Surgery, Deutsches Herzzentrum der Charité, 13353 Berlin, Germany.
The HeartMate 3 (HM3, Abbott) left ventricular assist device (LVAD) is the only commercially available option considered suitable for long-term circulatory support. External compression of the outflow graft causing obstruction (eOGO) is a serious adverse event affecting patients on long-term support. The obstruction occurs due to the accumulation of gelatinous substance between the bend relief and outflow graft.
View Article and Find Full Text PDFACS Chem Neurosci
November 2024
Department of Pharmaceutical Sciences, School of Pharmacy, College of Health and Human Sciences, North Dakota State University, Fargo, North Dakota 58108-6050, United States.
ASAIO J
July 2024
From the Division of Cardiothoracic Surgery, Department of Surgery, Medical College of Wisconsin, Milwaukee, Wisconsin.
As technology in mechanical circulatory support continues to advance, so does the utilization of durable left ventricular assist devices (LVADs). We present the first reported case of HeartMate 3 (HM3) LVAD bend relief disconnection causing laceration of the outflow graft. Our patient is a 34 year old male with end-stage heart failure status post-HM3 implantation with recent debridement of a driveline infection who presented with weakness and sanguineous drainage 5 days after an episode of heavy lifting associated with a palpable substernal pop.
View Article and Find Full Text PDFJ Heart Lung Transplant
November 2024
Division of Cardiovascular Medicine, University of Utah School of Medicine, Salt Lake City, Utah.
While notable improvements in survival, the incidence of hemocompatibility-related adverse events, hospitalizations, and cost have been demonstrated with the only commercially available durable left ventricular assist device, a category of pump malfunctions characterized by outflow graft obstruction has been noted with broader use and clinical follow-up of recipients of this technology. Of particular concern is the accumulation of acellular biodebris between the outflow graft and bend relief covering the outflow graft at its origin with the pump (which we term extrinsic outflow graft obstruction at the bend relief). This process tends to be insidious, occurs late in the postoperative course, can be challenging to diagnose, and can result in significant morbidity and mortality.