Purpose: Shaping and assembling contemporary external fixators rapidly for the severe mandibular fractures remains a challenge, especially in emergency circumstance. We designed a novel external fixator that incorporates universal joints to provide the stabilization for mandibular comminuted fractures. This study aims to confirm the efficacy of this novel external fixator through biomechanical tests in vitro and animal experiments.
Methods: In vitro biomechanical tests were conducted using 6 fresh canine with mandibular defect to simulate critical comminuted fractures. Three mandibles were stabilized by the novel external fixator and other mandibles were fixed by 2.5 mm reconstruction plates. All fixed mandibles were subjected to loads of 350 N on the anterior regions of teeth and 550 N on the first molar of the unaffected side. The stability was evaluated based on the maximum displacement and the slope of the load-displacement curve. In animal experiments, 9 beagles with comminuted mandibular fractures were divided into 3 groups, which were treated with the novel external fixation, reconstruction plate, and dental arch bar, respectively. The general observation, the changes in animals' weight, and the surgical duration were recorded and compared among 3 groups. The CT scans were performed at various intervals of 0 day (immediately after the surgery), 3 days, 7 days, 14 days, 21 days, and 28 days to analyze the displacement of feature points on the canine mandible and situation of fracture healing at 28 days. The statistical significance was assessed by the two-way analysis of variance test followed by the Bonferroni test, enabling multiple comparisons for all tests using GraphPad Prism10.1.0 (GraphPad Inc, USA).
Results: The outcomes of the biomechanical tests indicated that no statistically significant differences were found in terms of the maximum displacement (p = 0.496, 0.079) and the slope of load displacement curves (p = 0.374, 0.349) under 2 load modes between the external and internal fixation groups. The animal experiment data showed that there were minor displacements of feature points between the external and internal fixation groups without statistic difference, while the arch bar group demonstrated inferior stability. The CT analysis revealed that the best fracture healing happened in the internal fixation group, followed by the external fixation and arch baring at 28 days after fixation. The external fixation group had the shortest fixation duration (25.67 ± 3.79) min compared to internal fixation ((70.67 ± 4.51) min, p < 0.001) and arch baring ((42.00 ± 3.00) min, p = 0.046).
Conclusion: The conclusion of this study highlighted the efficacy and reliability of this novel external fixator in managing mandibular fractures rapidly, offering a viable option for the initial stabilization of comminuted mandibular fractures in the setting of emergency rescue.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11401486 | PMC |
http://dx.doi.org/10.1016/j.cjtee.2024.06.002 | DOI Listing |
Mater Today Bio
February 2025
Northern Jiangsu People's Hospital, Yangzhou, Jiangsu, China.
Spinal cord injury (SCI) is a neurological condition that causes significant loss of sensory, motor, and autonomic functions below the level of injury. Current clinical treatment strategies often fail to meet expectations. Hyaluronidase is typically associated with tumor progression and bacterial infections.
View Article and Find Full Text PDFMater Today Bio
February 2025
Department of Radiation Oncology, Radiation Oncology Key Laboratory of Sichuan Province, Sichuan Clinical Research Center for Cancer, Sichuan Cancer Hospital & Institute, Sichuan Cancer Center, Affiliated Cancer Hospital of University of Electronic Science and Technology of China, Chengdu, 610041, China.
Esophageal cancer is the eighth most common cancer worldwide and the sixth leading cause of cancer-related deaths. In this study, we propose a novel esophageal stent equipped with a wireless, battery-free, and movable photodynamic therapy (PDT) unit designed to treat esophageal tumors with flexibility, precision, and real-time control. This system integrates a PDT unit and an electrochemical pneumatic soft actuator into a conventional esophageal stent.
View Article and Find Full Text PDFCureus
December 2024
Neurological Surgery, Upstate University Hospital, Syracuse, USA.
Over the past two decades, despite the emergence of various novel therapies for glioblastoma, patient survival outcomes remain poor, particularly in the recurrent stage of the disease. Cesium-131 (Cs-131) brachytherapy presents a promising treatment option for patients with newly diagnosed and recurrent brain neoplasms, enabling the initiation of radiation therapy at the time of tumor resection. This approach eliminates the typical delay in therapy following surgery and the need for multiple return visits for fractionated external beam radiotherapy.
View Article and Find Full Text PDFCureus
December 2024
Department of Medicine, MetroWest Medical Center, Framingham, USA.
Localized inflammatory reactions in patients with past procedural history of intradermal injections can quickly drive the clinician's attention towards a diagnosis of soft-tissue infection in the context of symptoms such as fever, malaise, and local induration of the adipose panniculus. However, in patients with a long-term history of granulomatous events, a rheumatologic approach must be taken into consideration when the clinical course overwhelms the odds for more conventional diagnoses. In this case, a 39-year-old female patient who underwent bilateral lower limbs intradermal filllers presented with a two-year clinical course of repetitive flares of external bilateral hip tenderness, pain that limits her walking, soft-tissue nodular inflammation, redness, fever and a soft mobile nonpainful right supraclavicular lymphadenopathy.
View Article and Find Full Text PDFHeliyon
December 2024
Department of Pediatric and Adolescent surgery, University Hospital of Rouen, 76000, Rouen, France.
Objective: This study created and evaluated the face validity of a novel, low-cost, three-dimensional printed pediatric laparoscopic trainer and compared it to a standard laparoscopic trainer (LT).
Design: We developed the Space Child Laparoscopic Trainer (SCLT) using fused deposition modeling technology with a Bambulab P1S 3D printer.The cost per unit was five euros and the assembly time was 10 min.