BACKGROUND Pulsed field ablation (PFA), as a non-thermal ablation modality, has received increasing attention. The aim of this study was to evaluate the effect of PFA upon His bundle via its implementation with different voltages on the maximum His bundle potential in canines, providing scientific basis for clinical application. MATERIAL AND METHODS Pulsed electrical field energy was delivered from a ablation catheter to the maximum His potential of 7 dogs, followed by a series of electrogram and histology assessments. RESULTS The baseline AH and HV intervals were 55.3±3.7 ms (range, 53.0-59.0 ms), and 34.9±1.3 ms (range, 34.0-36.0 ms), respectively, which were elevated to 65.0±5.4 ms (range, 59.0-70.0 ms) and 35.7±2.7 ms (range, 34.0-37.0 ms) after PFA. Before ablation and immediately after the recovery of third-degree AVB, the AH interval was prolonged (P<0.05) while the HV interval remained unchanged (P>0.05). After ablation, all 7 canines experienced transient third-degree AVB, with a voltage-dependent duration. Masson staining results revealed no apparent damage in His bundle cells. CONCLUSIONS Within a certain voltage range of pulse electric field, ablation of the maximum His potential in canines can result in transient third-degree AVB, providing a new route for guiding safe ablation of para-Hisian arrhythmia.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11297367 | PMC |
http://dx.doi.org/10.12659/MSM.945007 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!