Background: The accuracy of infant intake using test weights (TWs), the change in weight before and after an infant feeds, has only been validated in hospitalized premature infants. This study's primary aim was to identify how accurate parent infants/toddler (< 2 years old) TWs are at measuring infant intake.

Methods: Data were collected from 101 paired bottle and infant/toddler TWs with 31 participants. Parents participated in the feeding sessions by completing infant/toddler TWs blinded to the researcher. Research assistants completed human milk bottle TWs. Infants were fed previously expressed human milk, initially 30 g, but volumes were increased to not exceed the scale's capacity.

Results: The mean difference between the bottle TWs measured using the Tanita and OHAUS scales was not significantly different from zero (95% CI (Tanita - OHAUS): (-0.251, 0.108) g). The mean difference between infant/toddler and bottle TWs was significantly different from 0 (95% CI (infant-bottle): (-3.45, -0.915 g or -3.57, -0.95 mL). Infant/toddler and bottle TWs were in agreement with a difference of 2.18 g (SD = 6.63) or 2.25 mL within the scale stated accuracy.

Conclusions: The Tanita infant digital scale accurately measures bottle TWs. The differences in parent infant/toddler TWs are within a clinically acceptable range.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11270915PMC
http://dx.doi.org/10.1186/s40795-024-00911-4DOI Listing

Publication Analysis

Top Keywords

test weights
8
comparison human
4
human milk
4
milk bottle
4
bottle infant/toddler
4
infant/toddler test
4
weights community
4
community setting
4
setting background
4
background accuracy
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!