Aims: Dried blood volumetric absorptive microsamples (VAMS) may facilitate home-based sampling to enhance therapeutic drug monitoring after transplantation. This study aimed to clinically validate a liquid chromatography-tandem mass spectrometry assay using 2 VAMS devices with different sampling locations (Tasso-M20 for the upper arm and Mitra for the finger). Patient preferences were also evaluated.
Methods: Clinical validation was performed for tacrolimus and mycophenolic acid by comparison of paired VAMS and venipuncture samples using Passing-Bablok regression and Bland-Altman analysis. Conversion of mycophenolic acid VAMS to serum concentrations was evaluated using haematocrit-dependent formulas and fixed correction factors defined a priori. Patients' perspectives, including useability, acceptability and feasibility, were also investigated using established questionnaires.
Results: Paired samples (n = 50) were collected from 25 kidney transplant recipients. Differences for tacrolimus whole-blood concentration were within ±20% for 86 and 88% of samples from the upper arm and fingerstick, respectively. Using correction factors of 1.3 for the upper-arm and 1.47 for finger-prick samples, 84 and 76% of the paired samples, respectively, were within ±20% for mycophenolic acid serum concentration. Patient experience surveys demonstrated limited pain and acceptable useability of the upper-arm device.
Conclusions: Tacrolimus and mycophenolic acid can be measured using 2 common VAMS devices with similar analytical performance. Patients are supportive of home-based monitoring with a preference for the Tasso-M20 device.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/bcp.16182 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!