Radiomic feature robustness in CT scans affected by fiducial marker induced streak artifacts for patients with hepatocellular carcinoma.

Med Phys

The D-Lab, Department of Precision Medicine, GROW School for Oncology and Reproduction, Maastricht University, Maastricht, The Netherlands.

Published: November 2024

Background: Stereotactic body radiation therapy for hepatocellular carcinoma necessitates the implantation of gold fiducial markers in the liver, resulting in artifacts on computed tomography (CT) scans, which affect radiomic feature values.

Purpose: This report aims to assess the effect of these artifacts on radiomic features and how removing CT slices affects radiomic features extracted from 3D regions of interest (ROI).

Methods: First, the range variation in 38 tumor contours unaffected by artifacts was assessed after sequentially and randomly removing 25%, 50%, 75% of slices. Subsequently, the agreement of feature values before and after removing ROI slices containing artifacts from 186 patients' CT scans was assessed with Lin's concordance correlation coefficient (CCC) and a Wilcoxon signed-rank test.

Results: In artifact-free tumor volumes, at least 71% of features remain robust with up to 50% of slices removed, while 56% remains robust with up to 75% of slices removed. When comparing contours before and after removing slices containing artifacts, around a third of features in the tumor contour and surrounding area remain robust (CCC > 0.9), compared to 44% in the healthy liver. Concerning the tumor, 13% (Gray Level Size Zone Matrix) to 61% (first order) of the features remain robust (CCC > 0.9). Over 90% of features differ significantly as assessed by Wilcoxon signed-rank test, however.

Conclusions: This study demonstrates that removing slices containing artifacts is a feasible solution for the CT fiducial problem in this patient population and provides insight into which features are affected.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.17329DOI Listing

Publication Analysis

Top Keywords

removing slices
12
slices artifacts
12
remain robust
12
radiomic feature
8
hepatocellular carcinoma
8
radiomic features
8
75% slices
8
wilcoxon signed-rank
8
features remain
8
slices removed
8

Similar Publications

Stored elastic bending tension as a mediator of embryonic body folding.

Cell Rep

January 2025

Department of Genetics and Developmental Biology, Rappaport Faculty of Medicine, Technion-Israel Institute of Technology, Haifa 31096, Israel. Electronic address:

During development, amniote vertebrate embryos transform from a flat sheet into a three-dimensional cylindrical form through ventral folding of the lateral sides of the sheet (the lateral plate [LP]) and their fusion in the ventral midline. Using a chick embryo slice system, we find that the flat stage is actually a poised balance of opposing dorsal and ventral elastic bending tensions. An intact extracellular matrix (ECM) is required for generating tension, as localized digestion of ECM dissipates tension, while removal of endoderm or ectoderm layers has no significant effect.

View Article and Find Full Text PDF

Step-by-step hysteroscopic treatment of FIGO type 3 myoma with the cold loop technique.

Fertil Steril

December 2024

Section of Obstetrics and Gynecology, Department of Medicine and Surgery, University of Perugia, Perugia, Italy. Electronic address:

Objective: To demonstrate the "cold loop technique" for the hysteroscopic treatment of FIGO type 3 myomas.

Design: Step-by-step demonstration of the technique using educative video.

Subject: A 45-year-old infertile patient with repeated oocyte donor IVF failures affected by a FIGO type 3 myoma.

View Article and Find Full Text PDF

Background: Prior studies have established that macroaggregated albumin (MAA)-SPECT/CT offers more robust lung shunt fraction (LSF) and lung mean absorbed dose (LMD) estimates in Y radioembolization in comparison to planar imaging. However, incomplete SPECT/CT coverage of the lungs is common due to clinical workflows, complicating its potential use for LSF and LMD calculations. In this work, lung truncation in MAA-SPECT/CT was addressed via correction strategies to improve Y treatment planning.

View Article and Find Full Text PDF
Article Synopsis
  • Event cameras mimic biological vision, providing high dynamic range and temporal resolution, which improve motion estimation in complex scenes, but they have limitations like low spatial resolution and data redundancy.
  • Current pre-processing methods for event streams struggle with challenges such as varying speeds and multiple subjects, leading to inefficiencies in data handling.
  • The proposed Asynchronous Spike Dynamic Metric and Slicing algorithm (ASDMS) and Adaptive Spatiotemporal Subject Surface Compensation algorithm (ASSSC) enhance event stream segmentation and image quality, while a new evaluation metric, Actual Performance Efficiency Discrepancy (APED), measures the effectiveness of these improvements against existing methods.
View Article and Find Full Text PDF

The dorsomedial hypothalamus (DMH) receives inputs from the preoptic area (POA), where ambient temperature mediates physiological adaptations of energy expenditure and food intake. Warm-activated POA neurons suppress energy expenditure via brown adipose tissue (BAT) projecting neurons in the dorsomedial hypothalamus/dorsal hypothalamic area (dDMH/DHA). Our earlier work identified leptin receptor (Lepr)-expressing, BAT-projecting dDMH/DHA neurons that mediate metabolic leptin effects.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!