Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: is an important bacterial pathogen, particularly as a cause of nosocomial infections in hospitalized patients. Only few reports exist in which cystoscopes were implicated as an outbreak source. We describe an investigation into the cause of a sudden increase in the number of urinary tract infections (UTI) with . in patients after cystoscopy. In addition, we share the lessons learned and measures taken to reduce the risk of similar infections in the future.
Presentation Of Case: Over a period of two weeks the urology outpatient department noticed a UTI in four patients following cystoscopy. An investigation was started for a common source of the outbreak in the urological treatment room. Additional screening of patients revealed a total of eleven males with UTI following cystoscopy. The infections were found to be due to a defective drying cabinet, which lacked an alarm signaling in case of loss of airflow. Amplified fragment length polymorphism (AFLP) analysis revealed that isolates from three patients and six isolates from environmental cultures (including cystoscopes from the drying cabinet) genotypically belonged to one strain.
Discussion: The AFLP results suggest that contaminated cystoscopes caused UTI in 11 patients, with the drying cabinet as site of transfer of the infective strain. To our knowledge, this is the first report describing a malfunctioning drying cabinet as source of an outbreak following cystoscopy.
Conclusion: In case of concomitant infections, cystoscopes and drying cabinets should be suspected as a potential source. Molecular techniques are helpful in investigating the epidemiology of an outbreak.