An Assessment of 1067 Light Curing Units Measured Twice Between 2015 and 2021.

Eur J Prosthodont Restor Dent

BlueLight Analytics, 2625 Joseph Howe Dr #24, Halifax, Nova Scotia B3L 4G4, Canada.

Published: September 2024

Introduction: The physical condition of the tip and the irradiance of LCUs used in dental offices in the United States and Canada was assessed twice.

Methods: The tip irradiance and physical condition of the tip of 1067 LCUs from 544 dental clinics were assessed twice between 2015 and 2021 using the CheckMARC (BlueLight) radiometer. The irradiance values were compared by paired Student's t-test (α = 0.05).

Results: There was a wide range of irradiance values (200 to 3,777 mW/cm²). At the first assessment, only 26% of the LCUs had no visible damage or debris on the LCU tip. When retested up to 36 months later, 70% of the LCUs had their tip damage or debris issues entirely or partially resolved. 2.6% of the LCUs delivered an irradiance below 500 mW/cm² on the first and 1.6% on the second assessments. For the 5 brands that had a sufficient sample size, it was concluded that the irradiance increased when the tip damage or presence of tip debris was resolved.

Conclusions: Initially, most LCUs (74%) had at least one problem related to damage or debris on the LCU tip. The irradiance increased when the tip damage or debris was resolved.

Download full-text PDF

Source
http://dx.doi.org/10.1922/EJPRD_2693Price10DOI Listing

Publication Analysis

Top Keywords

damage debris
16
2015 2021
8
physical condition
8
irradiance values
8
26% lcus
8
debris lcu
8
irradiance increased
8
increased damage
8
irradiance
7
lcus
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!