  • Transfemoral prosthesis users commonly experience falls due to stumbling, which pose a significant risk for injury, prompting this study to explore their recovery responses and identify flaws in current prosthetic designs.
  • The study involved six prosthesis users who were intentionally perturbed while walking on a treadmill to collect data on their recovery from falls, finding that more than half of the trials resulted in falls, with the highest occurrences during early swing.
  • Key deficiencies in prostheses were identified, including poor resistance to knee flexion and limited swing extension, which hinder effective recovery and stability for users.

Article Abstract

Background: Falls due to stumbling are prevalent for transfemoral prosthesis users and may lead to increased injury risk. This preliminary case series analyzes the transfemoral prosthesis user stumble recovery response to highlight key deficits in current commercially-available prostheses and proposes potential interventions to improve recovery outcomes.

Methods: Six transfemoral prosthesis users were perturbed on their prosthetic limb at least three times while walking on a treadmill using obstacle perturbations in early, mid and late swing. Kinematic data were collected to characterize the response, while fall rate and key kinematic recovery metrics were used to assess the quality of recovery and highlight functional deficits in current commercially-available prostheses.

Results: Across all participants, 13 (54%) of the 24 trials resulted in a fall (defined as > 50% body-weight support) with all but one participant (83%) falling at least once and two participants (33%) falling every time. In contrast, in a previous study of seven young, unimpaired, non-prosthesis users using the same experimental apparatus, no falls occurred across 190 trials. For the transfemoral prosthesis users, early swing had the highest rate of falling at 64%, followed by mid-swing at 57%, and then late swing at 33%. The trend in falls was mirrored by the kinematic recovery metrics (peak trunk angle, peak trunk angular velocity, forward reach of the perturbed limb, and knee angle at ground contact). In early swing all four metrics were deficient compared to non-prosthesis user controls. In mid swing, all but trunk angular velocity were deficient. In late swing only forward reach was deficient.

Conclusion: Based on the stumble recovery responses, four potential deficiencies were identified in the response of the knee prostheses: (1) insufficient resistance to stance knee flexion upon ground contact; (2) insufficient swing extension after a perturbation; (3) difficulty initiating swing flexion following a perturbation; and (4) excessive impedance against swing flexion in early swing preventing the potential utilization of the elevating strategy. Each of these issues can potentially be addressed by mechanical or mechatronic changes to prosthetic design to improve quality of recovery and reduce the likelihood a fall.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11245817PMC
http://dx.doi.org/10.1186/s12984-024-01402-0DOI Listing

Publication Analysis

Top Keywords

transfemoral prosthesis
20
prosthesis users
16
stumble recovery
12
late swing
12
early swing
12
swing
10
case series
8
recovery
8
recovery responses
8
deficits current
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!