Objective: To describe the accuracy of HealthVCF, a software product that uses artificial intelligence, in the detection of incidental moderate-to-severe vertebral compression fractures (VCFs) on chest and abdominal computed tomography scans.

Materials And Methods: We included a consecutive sample of 899 chest and abdominal computed tomography scans of patients 51-99 years of age. Scans were retrospectively evaluated by the software and by two specialists in musculoskeletal imaging for the presence of VCFs with vertebral body height loss > 25%. We compared the software analysis with that of a general radiologist, using the evaluation of the two specialists as the reference.

Results: The software showed a diagnostic accuracy of 89.6% (95% CI: 87.4-91.5%) for moderate-to-severe VCFs, with a sensitivity of 73.8%, a specificity of 92.7%, and a negative predictive value of 94.8%. Among the 145 positive scans detected by the software, the general radiologist failed to report the fractures in 62 (42.8%), and the algorithm detected additional fractures in 38 of those scans.

Conclusion: The software has good accuracy for the detection of moderate-to-severe VCFs, with high specificity, and can increase the opportunistic detection rate of VCFs by radiologists who do not specialize in musculoskeletal imaging.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11235064PMC
http://dx.doi.org/10.1590/0100-3984.2023.0102DOI Listing

Publication Analysis

Top Keywords

computed tomography
12
artificial intelligence
8
moderate-to-severe vertebral
8
vertebral compression
8
compression fractures
8
tomography scans
8
chest abdominal
8
abdominal computed
8
musculoskeletal imaging
8
general radiologist
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!