Purpose: Modular cementless knee arthroplasty systems are capable of precise reconstruction of the mechanical axis. However, they are considered more susceptible to complications. In contrast, non-modular cemented systems are said to be more forgiving and show good long-term results. The aim of this study was to investigate the resulting leg axis after implantation of a non-modular cemented rotating hinged knee prosthesis. Furthermore, potential risk factors for the occurrence of malalignment and complications should be identified.

Methods: Between 2005 and 2015, 115 patients could be included in this monocentric retrospective cohort study. All patients underwent primary hinged non-modular cemented total knee arthroplasty. Preoperative and postoperative standardized long radiographs were analysed to determine resulting leg axis. Furthermore, epidemiological and intraoperative data as well as perioperative complications were surveyed.

Results: Average leg axis was 5.8° varus preoperatively and 0.6° valgus postoperatively. Considering an axis deviation of 3° as the target corridor, 27% of all cases examined were outside the desired range. 21% cases showed a femoral deviation from the target corridor and 15% showed a tibial deviation. There was a significant relationship between the preoperative mLDFA and the mechanical alignment of the femoral component (R = 0.396, p < 0.001) as well as between the preoperative mMPTA and the mechanical alignment of the tibial component (R = 0.187, p = 0.045). The mean operative duration was 96 min. No periprosthetic fractures were observed within the study cohort.

Conclusion: The main result of the present work is that a non-modular cemented rotating hinged knee arthroplasty system can reconstruct the mechanical leg axis precisely and comparable to modular cementless and unconstrained total knee prostheses. Component malalignment is primarily dependent upon extraarticular deformity preoperatively. Periprosthetic fracture rates and duration of surgery were lower compared with current literature.

Level Of Evidence: Level III: Retrospective cohort study.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11417076PMC
http://dx.doi.org/10.1007/s00402-024-05409-zDOI Listing

Publication Analysis

Top Keywords

leg axis
16
non-modular cemented
16
knee arthroplasty
12
reconstruction mechanical
8
total knee
8
target corridor
8
axis
6
leg
4
mechanical leg
4
non-modular
4

Similar Publications

: A common problem in pediatric orthopedics is leg length discrepancy (LLD). In adulthood, this may result in overload and degenerative changes in the lumbar spine, hip, and knee joints of the longer limb, and the fixed equinus position of the foot of the shorter limb. Surgical treatment using temporary epiphysiodesis with eight-plate implants is a minimally invasive, safe, and patient-tolerated procedure in LLD.

View Article and Find Full Text PDF

Purpose Of The Study: Temporary hemiepiphyseodesis using figure-eight plates is currently one of the main surgical techniques to correct axial deformities of lower limbs in paediatric patients. Comprehensive analysis, correct indication and monitoring of treatment are the basic prerequisites for successful therapy. The aim of the study was to analyse parameters that could become an alternative to standard parameters used nowadays, namely the inserted screw angle (SA), and a new parameter - condylar ratio (CR).

View Article and Find Full Text PDF

Background: Anterior cruciate ligament (ACL) stress techniques-including single-leg stress radiographs, Telos, and KT-1000 arthrometer-are highly accessible and can provide additional diagnostic information to assess ACL and ACL graft integrity. The degree of anterior tibial translation (ATT) may be useful in guiding treatment when a diagnosis on magnetic resonance imaging is not conclusive or for judging if additional treatments, such as anterolateral complex augmentation, may be necessary.

Purpose/hypothesis: The purpose of this study was to evaluate the effect of increasing posterior tibial slope (PTS) on baseline tibial position (BTP) and side-to-side differences (SSD) in ATT.

View Article and Find Full Text PDF
Article Synopsis
  • Knee pain often involves varus deformity and unicompartmental osteoarthritis, where high tibial valgus osteotomy (HTO) has emerged as an effective treatment option to delay joint replacement.
  • A case report highlighted two patients with bilateral knee arthritis who underwent medial opening-wedge high tibial osteotomy using Dr. Saigal's plate, both achieving good range of motion and minimal complications post-surgery.
  • Comprehensive evaluations, including radiological assessments and specific surgical techniques like the Miniaci Method, were employed to ensure successful outcomes, with both patients showing significant improvement after one year.
View Article and Find Full Text PDF

Objectives: Although total hip arthroplasty (THA) effectively alleviates pain and restores joint function in the end-stage hip disease, challenges remain in achieving precise osteotomy and minimizing subjective dependency on prosthesis positioning. This study aims to evaluate the efficacy and safety of preoperative virtual planning and navigation templates compared to conventional techniques, providing new methods to enhance the precision and personalization of THA.

Methods: During the period from 2022 to 2023, we conducted a retrospective case-control study on 74 patients who underwent THA surgery at our hospital, based on the inclusion and exclusion criteria.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!