Technical note: Can photon-counting CT improve PET/CT's quantitative accuracy?

Med Phys

Department of Radiology and Radiological Science, Department of Nuclear Medicine, Medical University of South Carolina, Charleston, South Carolina, USA.

Published: October 2024

Background: Linear attenuation coefficients (LACs) in positron emission tomography combined with computed tomography (PET/CT) are derived from CT scans that utilize energy-integrating detectors (EID-CT). These LACs are inaccurate when iodine contrast has been injected. Photon counting detector CT (PCD-CT) may be able to improve the accuracy.

Purpose: To investigate whether PCD-CT can improve PET/CT quantitative accuracy.

Methods: Two experiments were performed: one with CT only and one that combined PET and CT. The first experiment used an electron density phantom whose inserts were imaged with EID-CT and PCD-CT. The inserts simulated normal human tissues, including bone and iodinated blood. In the case of PCD-CT, virtual-monoenergetic images at 190 keV were created. LACs were derived in each case and compared against known values. For inserts with iodine, more accurate LACs were expected with PCD-CT. The second experiment involved a custom PET phantom with various materials simulating human tissues (blood, iodinated blood, and bone) and F radioactivity. Data were first acquired with an EID-CT-based PET/CT system and then separately in a PCD-CT system without PET. PET images were reconstructed using LAC from EID-CT and PCD-CT. PET image values were compared against known activity values using recovery coefficients (RC).

Results: In the first experiment, LAC based on EID-CT were in error by as much as 18%, whereas the corresponding PCD-CT based measurements were within 3%. In the second experiment, minimum, maximum, and mean RC were (96.1%, 115.4%, and 103.8%) for the EID-CT method, and (95.8%, 105.5%, and 101.0%) for the PCD-CT method. The consistency of PET images in body and head orientations was improved.

Conclusions: PCD-CT can acquire the information needed for accurate LAC for PET reconstruction in a single spiral acquisition.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.17299DOI Listing

Publication Analysis

Top Keywords

pcd-ct
10
pcd-ct improve
8
eid-ct pcd-ct
8
human tissues
8
iodinated blood
8
second experiment
8
pet images
8
pet
7
eid-ct
5
technical note
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!