Knowledge of the minimal detectable bone fracture gap is essential in three-dimensional (3D) models, particularly in pre-operative planning of osteosynthesis to avoid overlooking gaps. In this study, defined incisions and bony displacements ranging from 100 to 400 µm were created in diaphyseal radii in 20 paired forearm specimens and verified with light microscopy. The specimens were scanned utilizing different computed tomography (CT) technologies/scanners, specimen positionings, scan protocols, image segmentations, and processing protocols. Inter- and intra-operator variabilities were reported as coefficient kappa. In CT images, fracture gaps of 100 µm and bone lamellae of 300 µm and 400 µm width were identified at a rate of 80 to 100%, respectively, independent of the investigated settings. In contrast, only 400µm incisions and bony displacements were visible in digital 3D models, with detection rates dependent on CT technology, image segmentation, and post-processing algorithm. 3D bone models based on state-of-the-art CT imaging can reliably visualize clinically relevant bone fracture gap sizes. However, verification of fractures to be surgically addressed should be verified with the original CT image series.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10278-024-01185-9DOI Listing

Publication Analysis

Top Keywords

bone fracture
12
minimal detectable
8
detectable bone
8
fracture gaps
8
fracture gap
8
incisions bony
8
bony displacements
8
400 µm
8
bone
5
fracture
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!