Aims And Objectives: To determine accuracy of pedicle screws placed by freehand, fluoroscopy-assistance and robotic-assistance with intraoperative image acquisition, and determine the presence of learning curve in robotic spine surgery in a prospective single centre study.
Materials And Methods: In a prospective study, a total of 1120 pedicle screws were placed in Freehand group (n = 175), 1250 screws were placed in fluoroscopy-assisted group (n = 172), and 1225 screws were inserted in Robotic-assisted group(n = 180). Surgical parameters and screw accuracy were analyzed between the three groups. The preoperative plan was overlapped with post operative O-arm scan to determine if the screws were executed as planned.
Results: The frequency of clinically acceptable screw placement (Gertzbein and Robbins grade A, B) in the Freehand, Fluoroscopy-assisted, and Robotic-assisted groups were 97.7 %, 98.6 %, and 99.34 % respectively. Higher pedicle screw accuracy, and lower blood loss were seen with robotic assistance. There was no significant difference in these parameters between surgeries commencing before and after 2 p.m. We found no statistically significant differences between the planned and executed screw trajectories in robotic assisted group irrespective of surgical experience.
Conclusion: The third-generation robotic-assisted pedicle screw placement system, used in conjunction with intraoperative 3D O-arm imaging, consistently lowered blood loss and increased accuracy of pedicle screw placement in the thoracolumbar spine. It also has easy adaptability into spine practice with minimal learning curve.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11225720 | PMC |
http://dx.doi.org/10.1016/j.jor.2024.06.011 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!