Background: A dual-chamber leadless pacemaker system has been designed for atrioventricular synchronous pacing using wireless, beat-to-beat, implant-to-implant (i2i) communication between distinct atrial and ventricular leadless pacemakers. The atrioventricular synchrony achieved across various ambulatory scenarios has yet to be systematically evaluated.
Methods: A prospective, single-arm, unblinded, multicenter, international clinical trial of the leadless pacemaker system was conducted in patients with a conventional dual-chamber pacing indication enrolled from February 2022 to March 2023. Leadless pacemaker systems were implanted, and 12-lead Holter electrocardiographic recordings were collected 3 months after implantation over various postures/activities: sitting, supine, left lateral recumbent, right lateral recumbent, standing, normal walk, and fast walk. An independent Holter core laboratory performed a manual adjudication of the percent of atrioventricular synchronous beats using the standard 300-millisecond PR interval limit. Atrium-to-ventricle and ventricle-to-atrium i2i communication success rates were also assessed. Post hoc summary statistics describing the relationships between atrioventricular synchrony and i2i success, posture/activity, implantation indication, atrioventricular event, and heart rate were calculated.
Results: In the evaluable population (n=384 of 464 enrolled [83%]; 61% male; age, 70 years; weight, 82 kg; 60% ejection fraction; 95% of beats evaluable), the mean atrioventricular synchrony of 98% of beats observed across all postures using the standard 300-millisecond limit was greater than both atrial-to-ventricular i2i (94%) and ventricular-to-atrial i2i (94%; <0.001), exceeding both i2i values in 95% of patients. Atrioventricular synchrony was achieved in >95% of evaluable beats across all postures/activities, implantation indications, atrioventricular paced/sensed event combinations, and heart rate ranges (including >100 bpm).
Conclusions: This dual-chamber leadless pacemaker system demonstrated atrioventricular synchrony in 98% of evaluable beats at 3 months after implantation. Atrioventricular synchrony was maintained across postures/activities and remained robust for heart rates >100 bpm.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11305627 | PMC |
http://dx.doi.org/10.1161/CIRCULATIONAHA.124.069006 | DOI Listing |
JACC Clin Electrophysiol
December 2024
Division of Cardiology, Department of Medicine, Weill Cornell Medicine, New York-Presbyterian Hospital, New York, New York, USA. Electronic address:
JACC Case Rep
November 2024
Division of Cardiovascular Medicine, Hillel Yaffe Medical Center, The Ruth and Bruce Rappaport Faculty of Medicine, Technion, Haifa, Israel.
We herein describe a case of successful atrioventricular synchrony at an 8-week follow-up in a patient who received a leadless pacemaker for recurrent right ventricular lead failures and had a pre-existing atrial transvenous pacemaker. Given the significant hemodynamic improvements, careful initial programming and adjustments during follow-up are needed.
View Article and Find Full Text PDFPacing Clin Electrophysiol
January 2025
Department of Cardiology, Chutoen General Medical Center, Kakegawa, Shizuoka, Japan.
Introduction: The VDD-leadless pacemaker aims to maintain high atrioventricular (AV) synchrony, but proper sensing setup is crucial.
Methods And Results: This case was an elderly woman with an AV block who received a dual-chamber pacemaker. Due to ventricular pacing failure, a Micra AV2 was implanted.
World J Cardiol
November 2024
Department of Cardiology, Dayanand Medical College and Hospital, Ludhiana 141001, Punjab, India.
The deleterious effects of long term right ventricular pacing are increasingly being recognized today. Current clinical practice favors the implantation of dual-chamber permanent pacemaker which maintains atrioventricular synchrony and is associated with better quality of life. However, despite the popular belief and common sense surrounding the superiority of dual-chamber pacing over single chamber pacing, the same has never been conclusively verified in clinical trials.
View Article and Find Full Text PDFArrhythm Electrophysiol Rev
November 2024
University of California Los Angeles (UCLA) Cardiac Arrhythmia Center Los Angeles, CA, US.
Permanent cardiac pacing in children with congenital complete atrioventricular block (CCAVB) and/or congenital heart disease (CHD) is challenging. Conduction system pacing (CSP) represents a novel pacing strategy aiming to preserve physiological ventricular activation. Patients with CCAVB or CHD are at high risk of developing pacing-induced cardiomyopathy with chronic conventional right ventricular myocardial pacing.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!