Background: Hospital articles processed by steam are widely used in the Central Sterile Supply Department (CSSD), responsible for due sterilization. Steam sterilization is discussed worldwide, aiming to protect patients. If steam is outside the specified requirements, the sterilization process may fail, resulting in the wet packs at the end of the sterilization cycle.

Methods: The present study evaluated the steam quality at Santa Catarina Hospital (São Paulo, SP, Brazil) from 2016 to 2022. Saturated steam containing noncondensable gases, excess condensate, or even superheat was characterized using the methodology indicated in the European Standard EN 285:2015.

Results: From 2016 to 2020, qualification tests showed that the saturated steam quality does not achieve standard limit parameters. Infrastructural maintenance actions were taken to adjust the saturated steam quality. In 2021, the steam quality followed technical standards, and its adequacy was confirmed in 2022.

Conclusions: The points developed by the hospital's maintenance department, the adoption of appropriate devices for this purpose, and the correct preventive maintenance in the autoclaves, together with the correct qualification of the equipment and proof of the steam quality, contributed to improve the safety of the hospital sterilization process and reduce the incidence of wet packages.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajic.2024.07.002DOI Listing

Publication Analysis

Top Keywords

steam quality
24
saturated steam
12
steam
10
wet packs
8
packs sterilization
8
sterilization process
8
sterilization
6
quality
5
quality monitoring
4
monitoring strategy
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!